1 |
|
---|
2 |
|
---|
3 | In this section, we describe the tests performed using light pulses of different colour,
|
---|
4 | pulse shapes and intensities with the MAGIC LED Calibration Pulser Box \cite{hardware-manual}.
|
---|
5 | \par
|
---|
6 | The LED pulser system is able to provide fast light pulses of 3--4\,ns FWHM
|
---|
7 | with intensities ranging from 3--4 to more than 500 photo-electrons in one inner photo-multiplier of the
|
---|
8 | camera. These pulses can be produced in three colours {\textit {\bf green, blue}} and
|
---|
9 | {\textit{\bf UV}}.
|
---|
10 |
|
---|
11 | \begin{table}[htp]
|
---|
12 | \centering
|
---|
13 | \begin{tabular}{|c|c|c|c|c|c|c|}
|
---|
14 | \hline
|
---|
15 | \hline
|
---|
16 | \multicolumn{7}{|c|}{The possible pulsed light colours} \\
|
---|
17 | \hline
|
---|
18 | \hline
|
---|
19 | Colour & Wavelength & Spectral Width & Min. Nr. & Max. Nr. & Secondary & FWHM \\
|
---|
20 | & [nm] & [nm] & Phe's & Phe's & Pulses & Pulse [ns]\\
|
---|
21 | \hline
|
---|
22 | Green & 520 & 40 & 6 & 120 & yes & 3--4 \\
|
---|
23 | \hline
|
---|
24 | Blue & 460 & 30 & 6 & 500 & yes & 3--4 \\
|
---|
25 | \hline
|
---|
26 | UV & 375 & 12 & 3 & 50 & no & 2--3 \\
|
---|
27 | \hline
|
---|
28 | \hline
|
---|
29 | \end{tabular}
|
---|
30 | \caption{The pulser colours available from the calibration system}
|
---|
31 | \label{tab:pulsercolours}
|
---|
32 | \end{table}
|
---|
33 |
|
---|
34 | Table~\ref{tab:pulsercolours} lists the available colours and intensities and
|
---|
35 | figures~\ref{fig:pulseexample1leduv} and~\ref{fig:pulseexample23ledblue} show exemplary pulses
|
---|
36 | as registered by the FADCs.
|
---|
37 | Whereas the UV-pulse is very stable, the green and blue pulses show sometimes smaller secondary
|
---|
38 | pulses after about 10--40\,ns from the main pulse.
|
---|
39 | One can see that the very stable UV-pulses are unfortunately only available in such intensities as to
|
---|
40 | not saturate the high-gain readout channel. However, the brightest combination of light pulses easily
|
---|
41 | saturates all channels in the camera, but does not reach a saturation of the low-gain readout.
|
---|
42 | \par
|
---|
43 | Our tests can be classified into three subsections:
|
---|
44 |
|
---|
45 | \begin{enumerate}
|
---|
46 | \item Un-calibrated pixels and events: These tests measure the percentage of failures of the extractor
|
---|
47 | resulting either in a pixel declared as un-calibrated or in an event which produces a signal ouside
|
---|
48 | of the expected Gaussian distribution.
|
---|
49 | \item Number of photo-electrons: These tests measure the reconstructed numbers of photo-electrons, their
|
---|
50 | spread over the camera and the ratio of the obtained mean values for outer and inner pixels, respectively.
|
---|
51 | \item Linearity tests: These tests measure the linearity of the extractor with respect to pulses of
|
---|
52 | different intensity and colour.
|
---|
53 | \item Time resolution: These tests show the time resolution and stability obtained with different
|
---|
54 | intensities and colours.
|
---|
55 | \end{enumerate}
|
---|
56 |
|
---|
57 | \begin{figure}[htp]
|
---|
58 | \centering
|
---|
59 | \includegraphics[width=0.48\linewidth]{1LedUV_Pulse_Inner.eps}
|
---|
60 | \includegraphics[width=0.48\linewidth]{1LedUV_Pulse_Outer.eps}
|
---|
61 | \caption{Example of a calibration pulse from the lowest available intensity (1\,Led UV).
|
---|
62 | The left plot shows the signal obtained in an inner pixel, the right one the signal in an outer pixel.
|
---|
63 | Note that the pulse height fluctuates much more than suggested from these pictures. Especially, a
|
---|
64 | zero-pulse is also possible.}
|
---|
65 | \label{fig:pulseexample1leduv}
|
---|
66 | \end{figure}
|
---|
67 |
|
---|
68 | \begin{figure}[htp]
|
---|
69 | \centering
|
---|
70 | \includegraphics[width=0.48\linewidth]{23LedsBlue_Pulse_Inner.eps}
|
---|
71 | \includegraphics[width=0.48\linewidth]{23LedsBlue_Pulse_Outer.eps}
|
---|
72 | \caption{Example of a calibration pulse from the highest available mono-chromatic intensity (23\,Leds Blue).
|
---|
73 | The left plot shows the signal obtained in an inner pixel, the right one the signal in an outer pixel.
|
---|
74 | One the left side of both plots, the (saturated) high-gain channel is visible,
|
---|
75 | on the right side from FADC slice 18 on,
|
---|
76 | the delayed low-gain
|
---|
77 | pulse appears. Note that in the left plot, there is a secondary pulses visible in the tail of the
|
---|
78 | high-gain pulse. }
|
---|
79 | \label{fig:pulseexample23ledblue}
|
---|
80 | \end{figure}
|
---|
81 |
|
---|
82 | We used data taken on the 7$^{th}$ of June, 2004 with different pulser LED combinations, each taken with
|
---|
83 | 16384 events. 19 different calibration configurations have been tested.
|
---|
84 | The corresponding MAGIC data run numbers range from nr. 31741 to 31772. These data was taken
|
---|
85 | before the latest camera repair access which resulted in a replacement of about 2\% of the pixels known to be
|
---|
86 | mal-functionning at that time.
|
---|
87 | There is thus a lower limit to the number of un-calibrated pixels of about 1.5--2\% of known
|
---|
88 | mal-functionning photo-multipliers.
|
---|
89 | \par
|
---|
90 | Although we had looked at and tested all colour and extractor combinations resulting from these data,
|
---|
91 | we refrain ourselves to show here only exemplary behaviour and results of extractors.
|
---|
92 | All plots, including those which are not displayed in this TDAS, can be retrieved from the following
|
---|
93 | locations:
|
---|
94 |
|
---|
95 | \begin{verbatim}
|
---|
96 | http://www.magic.ifae.es/~markus/pheplots/
|
---|
97 | http://www.magic.ifae.es/~markus/timeplots/
|
---|
98 | \end{verbatim}
|
---|
99 |
|
---|
100 | %%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
|
---|
101 |
|
---|
102 | \subsection{Un-Calibrated Pixels and Events}
|
---|
103 |
|
---|
104 | The MAGIC calibration software incorporates a series of checks to sort out mal-functionning pixels.
|
---|
105 | Except for the software bug searching criteria, the following exclusion criteria can apply:
|
---|
106 |
|
---|
107 | \begin{enumerate}
|
---|
108 | \item The reconstructed mean signal is less than 2.5 times the extractor resolution $R$ from zero.
|
---|
109 | (2.5 Pedestal RMS in the case of the simple fixed window extractors, see section~\ref{sec:pedestals}).
|
---|
110 | This criterium essentially cuts out
|
---|
111 | dead pixels.
|
---|
112 | \item The reconstructed mean signal error is smaller than its value. This criterium cuts out
|
---|
113 | signal distributions which fluctuate so much that their RMS is bigger than its mean value. This
|
---|
114 | criterium cuts out ``ringing'' pixels or mal-functionning extractors.
|
---|
115 | \item The reconstructed mean number of photo-electrons lies 4.5 sigma outside
|
---|
116 | the distribution of photo-electrons obtained with the inner or outer pixels in the camera, respectively.
|
---|
117 | This criterium cuts out pixels channels with apparently deviating (hardware) behaviour compared to
|
---|
118 | the rest of the camera readout\footnote{This criteria is not applied any more in the standard analysis,
|
---|
119 | although here, we kept using it}.
|
---|
120 | \item All pixels with reconstructed negative mean signal or with a
|
---|
121 | mean numbers of photo-electrons smaller than one. Pixels with a negative pedestal RMS subtracted
|
---|
122 | sigma occur, especially when stars are focussed onto that pixel during the pedestal taking (resulting
|
---|
123 | in a large pedestal RMS), but have moved to another pixel during the calibration run. In this case, the
|
---|
124 | number of photo-electrons would result artificially negative. If these
|
---|
125 | channels do not show any other deviating behaviour, their number of photo-electrons gets replaced by the
|
---|
126 | mean number of photo-electrons in the camera, and the channel is further calibrated as normal.
|
---|
127 | \end{enumerate}
|
---|
128 |
|
---|
129 | Moreover, the number of events are counted which have been reconstructed outside a 5 sigma region
|
---|
130 | from the mean signal. These events are called ``outliers''. Figure~\ref{fig:outlier} shows a typical
|
---|
131 | outlier obtained with the digital filter applied to a low-gain signal and figure~\ref{fig:unsuited:all}
|
---|
132 | shows the average number of all excluded pixels and outliers obtained from all 19 calibration configurations.
|
---|
133 | One can already see that the largest window sizes yield a high number of un-calibrated pixels, mostly
|
---|
134 | due to the missing ability to recognize the low-intensity pulses (see later). One can also see that
|
---|
135 | the amplitude extracting spline yields a higher number of outliers than the rest of the extractors.
|
---|
136 | The global champion in lowest number of un-calibrated pixels results to be
|
---|
137 | {\textit{\bf MExtractTimeAndChargeDigitalFilter}} with the correct calibration weights over 4 FADC slices
|
---|
138 | (extractor \#31). The one with the lowest number of outliers is
|
---|
139 | {\textit{\bf MExtractFixedWindowPeakSearch}} with an extraction range of 2 slices (extractor \#11).
|
---|
140 |
|
---|
141 | \begin{figure}[htp]
|
---|
142 | \centering
|
---|
143 | \includegraphics[width=0.95\linewidth]{Outlier.eps}
|
---|
144 | \caption{Example of an event classified as ``un-calibrated''. The histogram has been obtained
|
---|
145 | using the digital filter (extractor \#32) applied to a high-intensity blue pulse (run 31772).
|
---|
146 | The event marked as ``outlier'' clearly has been mis-reconstructed. It lies outside the 5 sigma
|
---|
147 | region from the fitted mean.}
|
---|
148 | \label{fig:outlier}
|
---|
149 | \end{figure}
|
---|
150 |
|
---|
151 | \begin{figure}[htp]
|
---|
152 | \centering
|
---|
153 | \includegraphics[height=0.75\textheight]{UnsuitVsExtractor-all.eps}
|
---|
154 | \caption{Uncalibrated pixels and outlier events averaged over all available
|
---|
155 | calibration runs.}
|
---|
156 | \label{fig:unsuited:all}
|
---|
157 | \end{figure}
|
---|
158 |
|
---|
159 | The following figures~\ref{fig:unsuited:5ledsuv},~\ref{fig:unsuited:1leduv},~\ref{fig:unsuited:2ledsgreen}
|
---|
160 | and~\ref{fig:unsuited:23ledsblue} show the resulting numbers of un-calibrated pixels and events for
|
---|
161 | different colours and intensities. Because there is a strong anti-correlation between the number of
|
---|
162 | excluded channels and the number of outliers per event, we have chosen to show these numbers together.
|
---|
163 |
|
---|
164 | \par
|
---|
165 |
|
---|
166 | \begin{figure}[htp]
|
---|
167 | \centering
|
---|
168 | \includegraphics[height=0.95\textheight]{UnsuitVsExtractor-5LedsUV-Colour-13.eps}
|
---|
169 | \caption{Uncalibrated pixels and outlier events for a typical calibration
|
---|
170 | pulse of UV-light which does not saturate the high-gain readout.}
|
---|
171 | \label{fig:unsuited:5ledsuv}
|
---|
172 | \end{figure}
|
---|
173 |
|
---|
174 | \begin{figure}[htp]
|
---|
175 | \centering
|
---|
176 | \includegraphics[height=0.95\textheight]{UnsuitVsExtractor-1LedUV-Colour-04.eps}
|
---|
177 | \caption{Uncalibrated pixels and outlier events for a very low
|
---|
178 | intensity pulse.}
|
---|
179 | \label{fig:unsuited:1leduv}
|
---|
180 | \end{figure}
|
---|
181 |
|
---|
182 | \begin{figure}[htp]
|
---|
183 | \centering
|
---|
184 | \includegraphics[height=0.95\textheight]{UnsuitVsExtractor-2LedsGreen-Colour-02.eps}
|
---|
185 | \caption{Uncalibrated pixels and outlier events for a typical green pulse.}
|
---|
186 | \label{fig:unsuited:2ledsgreen}
|
---|
187 | \end{figure}
|
---|
188 |
|
---|
189 | \begin{figure}[htp]
|
---|
190 | \centering
|
---|
191 | \includegraphics[height=0.95\textheight]{UnsuitVsExtractor-23LedsBlue-Colour-00.eps}
|
---|
192 | \caption{Uncalibrated pixels and outlier events for a high-intensity blue pulse.}
|
---|
193 | \label{fig:unsuited:23ledsblue}
|
---|
194 | \end{figure}
|
---|
195 |
|
---|
196 | One can see that in general, big extraction windows raise the
|
---|
197 | number of un-calibrated pixels and are thus less stable. Especially for the very low-intensity
|
---|
198 | \textit{\bf 1Led\,UV}-pulse, the big extraction windows summing 8 or more slices, cannot calibrate more
|
---|
199 | than 50\%
|
---|
200 | of the inner pixels (fig.~\ref{fig:unsuited:1leduv}). This is an expected behavior since big windows
|
---|
201 | add up more noise which in turn makes the search for the small signal more difficult.
|
---|
202 | \par
|
---|
203 | In general, one can also find that all ``sliding window''-algorithms (extractors \#17-32) discard
|
---|
204 | less pixels than the corresponding ``fixed window''-ones (extractors \#1--16). The digital filter with
|
---|
205 | the correct weights (extractors \#30-33) discards the least number of pixels and is also robust against
|
---|
206 | slight modifications of its weights (extractors \#28--30). The robustness gets lost when the high-gain and
|
---|
207 | low-gain weights are inverted (extractors \#31--39, see fig.~\ref{fig:unsuited:23ledsblue}).
|
---|
208 | \par
|
---|
209 | Also the ``spline'' algorithms on small
|
---|
210 | windows (extractors \#23--25) discard less pixels than the previous extractors.
|
---|
211 | \par
|
---|
212 | It seems also that the spline algorithm extracting the amplitude of the signal produces an over-proportional
|
---|
213 | number of excluded events in the low-gain. The same, however in a less significant manner, holds for
|
---|
214 | the digital filter with high-low-gain inverted weights. The limit of stability with respect to
|
---|
215 | changes in the pulse form seems to be reached, there.
|
---|
216 | \par
|
---|
217 | Concerning the numbers of outliers, one can conclude that in general, the numbers are very low never exceeding
|
---|
218 | 0.1\% except for the ampltiude-extracting spline which seems to mis-reconstruct a certain type of events.
|
---|
219 | \par
|
---|
220 | In conclusion, already this first test excludes all extractors with too large window sizes because
|
---|
221 | they are not able to extract cleanly small signals produced by about 4 photo-electrons. Moreover,
|
---|
222 | some extractors do not reproduce the signals as expected in the low-gain.
|
---|
223 |
|
---|
224 | %The excluded extractors are:
|
---|
225 | %\begin{itemize}
|
---|
226 | %\item: MExtractFixedWindow Nr. 3--5
|
---|
227 | %\item: MExtractFixedWindowSpline Nr. 6--11 (all)
|
---|
228 | %\item: MExtractFixedWindowPeakSearch Nr. 14--16
|
---|
229 | %\item: MExtractTimeAndChargeSlidingWindow Nr. 21--22
|
---|
230 | %\item: MExtractTimeAndChargeSpline Nr. 23 and 27
|
---|
231 | %\end{itemize}
|
---|
232 |
|
---|
233 | \clearpage
|
---|
234 |
|
---|
235 | %%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
|
---|
236 |
|
---|
237 | \subsection{Number of Photo-Electrons \label{sec:photo-electrons}}
|
---|
238 |
|
---|
239 | Assuming that the readout chain adds only negligible noise to the one
|
---|
240 | introduced by the photo-multiplier itself, one can make the assumption that the variance of the
|
---|
241 | true signal $S$ is the amplified Poisson variance of the number of photo-electrons,
|
---|
242 | multiplied with the excess noise of the photo-multiplier which itself is
|
---|
243 | characterized by the excess-noise factor $F$.
|
---|
244 |
|
---|
245 | \begin{equation}
|
---|
246 | Var(S) = F^2 \cdot Var(N_{phe}) \cdot \frac{<S>^2}{<N_{phe}>^2}
|
---|
247 | \label{eq:excessnoise}
|
---|
248 | \end{equation}
|
---|
249 |
|
---|
250 | After introducing the effect of the night-sky background (eq.~\ref{eq:rmssubtraction})
|
---|
251 | in formula~\ref{eq:excessnoise} and assuming that the variance of the number of photo-electrons is equal
|
---|
252 | to the mean number of photo-electrons (because of the Poisson distribution),
|
---|
253 | one obtains an expression to retrieve the mean number of photo-electrons impinging on the pixel from the
|
---|
254 | mean extracted signal $<\widehat{S}>$,
|
---|
255 | its variance $Var(\widehat{S})$ and the RMS of the extracted signal obtained from
|
---|
256 | pure pedestal runs $R$ (see section~\ref{sec:ffactor}):
|
---|
257 |
|
---|
258 | \begin{equation}
|
---|
259 | <N_{phe}> \approx F^2 \cdot \frac{<\widehat{S}>^2}{Var(\widehat{S}) - R^2}
|
---|
260 | \label{eq:pheffactor}
|
---|
261 | \end{equation}
|
---|
262 |
|
---|
263 | In theory, eq.~\ref{eq:pheffactor} must not depend on the extractor! Effectively, we will use it to test the
|
---|
264 | quality of our extractors by requiring that a valid extractor yields the same number of photo-electrons
|
---|
265 | for all pixels of a same type and does not deviate from the number obtained with other extractors.
|
---|
266 | As the camera is flat-fielded, but the number of photo-electrons impinging on an inner and an outer pixel is
|
---|
267 | different, we also use the ratio of the mean numbers of photo-electrons from the outer pixels to the one
|
---|
268 | obtained from the inner pixels as a test variable. In the ideal case, it should always yield its central
|
---|
269 | value of about 2.6$\pm$0.1~\cite{michele-diploma}.
|
---|
270 | \par
|
---|
271 | In our case, there is an additional complication due to the fact that the green and blue coloured light pulses
|
---|
272 | show secondary pulses which destroy the Poisson behaviour of the number of photo-electrons. We will
|
---|
273 | have to split our sample of extractors into those being affected by the secondary pulses and those
|
---|
274 | being immune to this effect.
|
---|
275 | \par
|
---|
276 | Figures~\ref{fig:phe:5ledsuv},~\ref{fig:phe:1leduv},~\ref{fig:phe:2ledsgreen}~and~\ref{fig:phe:23ledsblue} show
|
---|
277 | some of the obtained results. Although one can see a rather good stability for the standard
|
---|
278 | {\textit{\bf 5\,Leds\,UV}}\ pulse, except for the extractors {\textit{\bf MExtractFixedWindowPeakSearch}}, initialized
|
---|
279 | with an extraction window of 2 slices and {\textit{\bf MExtractTimeAndChargeDigitalFilter}}, initialized with
|
---|
280 | an extraction window of 4 slices (extractor \#29).
|
---|
281 | \par
|
---|
282 | There is a considerable difference for all shown non-standard pulses. Especially the pulses from green
|
---|
283 | and blue LEDs
|
---|
284 | show a clear dependency of the number of photo-electrons on the extraction window. Only the largest
|
---|
285 | extraction windows seem to catch the entire range of (jittering) secondary pulses and get the ratio
|
---|
286 | of outer vs. inner pixels right. However, they (obviously) over-estimate the number of photo-electrons
|
---|
287 | in the primary pulse.
|
---|
288 | \par
|
---|
289 | The strongest discrepancy is observed in the low-gain extraction (fig.~\ref{fig:phe:23ledsblue}) where all
|
---|
290 | fixed window extractors with too small extraction windows fail to reconstruct the correct numbers.
|
---|
291 | This has to do with the fact that
|
---|
292 | the fixed window extractors fail to do catch a significant part of the (larger) pulse because of the
|
---|
293 | 1~FADC slice event-to-event jitter.
|
---|
294 |
|
---|
295 |
|
---|
296 | \begin{figure}[htp]
|
---|
297 | \centering
|
---|
298 | \includegraphics[height=0.92\textheight]{PheVsExtractor-5LedsUV-Colour-13.eps}
|
---|
299 | \caption{Number of photo-electrons from a typical, not saturating calibration pulse of colour UV,
|
---|
300 | reconstructed with each of the tested signal extractors.
|
---|
301 | The first plots shows the number of photo-electrons obtained for the inner pixels, the second one
|
---|
302 | for the outer pixels and the third shows the ratio of the mean number of photo-electrons for the
|
---|
303 | outer pixels divided by the mean number of photo-electrons for the inner pixels. Points
|
---|
304 | denote the mean of all not-excluded pixels, the error bars their RMS.}
|
---|
305 | \label{fig:phe:5ledsuv}
|
---|
306 | \end{figure}
|
---|
307 |
|
---|
308 | \begin{figure}[htp]
|
---|
309 | \centering
|
---|
310 | \includegraphics[height=0.92\textheight]{PheVsExtractor-1LedUV-Colour-04.eps}
|
---|
311 | \caption{Number of photo-electrons from a typical, very low-intensity calibration pulse of colour UV,
|
---|
312 | reconstructed with each of the tested signal extractors.
|
---|
313 | The first plots shows the number of photo-electrons obtained for the inner pixels, the second one
|
---|
314 | for the outer pixels and the third shows the ratio of the mean number of photo-electrons for the
|
---|
315 | outer pixels divided by the mean number of photo-electrons for the inner pixels. Points
|
---|
316 | denote the mean of all not-excluded pixels, the error bars their RMS.}
|
---|
317 | \label{fig:phe:1leduv}
|
---|
318 | \end{figure}
|
---|
319 |
|
---|
320 | \begin{figure}[htp]
|
---|
321 | \centering
|
---|
322 | \includegraphics[height=0.92\textheight]{PheVsExtractor-2LedsGreen-Colour-02.eps}
|
---|
323 | \caption{Number of photo-electrons from a typical, not saturating calibration pulse of colour green,
|
---|
324 | reconstructed with each of the tested signal extractors.
|
---|
325 | The first plots shows the number of photo-electrons obtained for the inner pixels, the second one
|
---|
326 | for the outer pixels and the third shows the ratio of the mean number of photo-electrons for the
|
---|
327 | outer pixels divided by the mean number of photo-electrons for the inner pixels. Points
|
---|
328 | denote the mean of all not-excluded pixels, the error bars their RMS.}
|
---|
329 | \label{fig:phe:2ledsgreen}
|
---|
330 | \end{figure}
|
---|
331 |
|
---|
332 |
|
---|
333 | \begin{figure}[htp]
|
---|
334 | \centering
|
---|
335 | \includegraphics[height=0.92\textheight]{PheVsExtractor-23LedsBlue-Colour-00.eps}
|
---|
336 | \caption{Number of photo-electrons from a typical, high-gain saturating calibration pulse of colour blue,
|
---|
337 | reconstructed with each of the tested signal extractors.
|
---|
338 | The first plots shows the number of photo-electrons obtained for the inner pixels, the second one
|
---|
339 | for the outer pixels and the third shows the ratio of the mean number of photo-electrons for the
|
---|
340 | outer pixels divided by the mean number of photo-electrons for the inner pixels. Points
|
---|
341 | denote the mean of all not-excluded pixels, the error bars their RMS.}
|
---|
342 | \label{fig:phe:23ledsblue}
|
---|
343 | \end{figure}
|
---|
344 |
|
---|
345 | One can see that all extractors using a large window belong to the class of extractors being affected
|
---|
346 | by the secondary pulses, except for the digital filter. The only exception to this rule is the digital filter
|
---|
347 | which - despite of its 6 slices extraction window - seems to filter out all the secondary pulses.
|
---|
348 | \par
|
---|
349 | The extractor {\textit{\bf MExtractFixedWindowPeakSearch}} at low extraction windows apparently yields chronically low
|
---|
350 | numbers of photo-electrons. This is due to the fact that the decision to fix the extraction window is
|
---|
351 | made sometimes by an inner pixel and sometimes by an outer one since the camera is flat-fielded and the
|
---|
352 | pixel carrying the largest non-saturated peak-search window is more or less found by a random signal
|
---|
353 | fluctuation. However, inner and outer pixels have a systematic offset of about 0.5 to 1 FADC slices.
|
---|
354 | Thus, the extraction fluctuates artificially for one given channel which results in a systematically
|
---|
355 | large variance and thus in a systematically low reconstructed number of photo-electrons. This test thus
|
---|
356 | excludes the extractors \#11--13.
|
---|
357 | \par
|
---|
358 | Moreover, one can see that the extractors applying a small fixed window do not get the ratio of
|
---|
359 | photo-electrons correctly between outer to inner pixels for the green and blue pulses.
|
---|
360 | \par
|
---|
361 | The extractor {\textit{\bf MExtractTimeAndChargeDigitalFilter}} seems to be stable against modifications in the
|
---|
362 | exact form of the weights in the high-gain readout channel since all applied weights yield about
|
---|
363 | the same number of photo-electrons and the same ratio of outer vs. inner pixels. This statement does not
|
---|
364 | hold any more for the low-gain, as can be seen in figure~\ref{fig:phe:23ledsblue}. There, the application
|
---|
365 | of high-gain weights to the low-gain signal (extractors \#34--39) produces a too low number of photo-electrons
|
---|
366 | and also a too low ratio of outer vs. inner pixels.
|
---|
367 | \par
|
---|
368 | All sliding window and spline algorithms yield a stable ratio of outer vs. inner pixels in the low-gain,
|
---|
369 | however the effect of raising the number of photo-electrons with the extraction window is very pronounced.
|
---|
370 | Note that in figure~\ref{fig:phe:23ledsblue}, the number of photo-electrons rises by about a factor 1.4,
|
---|
371 | which is slightly higher than in the case of the high-gain channel (figure~\ref{fig:phe:2ledsgreen}).
|
---|
372 | \par
|
---|
373 | Concluding, there is no fixed window extractor yielding the correct number of photo-electrons
|
---|
374 | for the low-gain, except for the largest extraction window of 8 and 10 low-gain slices.
|
---|
375 | Either the number of photo-electrons itself is wrong or the ratio of outer vs. inner pixels is
|
---|
376 | not correct. All sliding window algorithms seem to reproduce the correct numbers if one takes into
|
---|
377 | account the after-pulse behaviour of the light pulser itself. The digital filter seems to be
|
---|
378 | unstable against exchanging the pulse form to match the slimmer high-gain pulses, though.
|
---|
379 |
|
---|
380 | \par
|
---|
381 | \ldots {\textit{\bf EXCLUDED : CW4, UV4 No stability High-gain vs. LoGain}}
|
---|
382 | \par
|
---|
383 |
|
---|
384 | %%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
|
---|
385 |
|
---|
386 | \subsection{Linearity \label{sec:calibration:linearity}}
|
---|
387 |
|
---|
388 | \begin{figure}[htp]
|
---|
389 | \centering
|
---|
390 | \includegraphics[width=0.99\linewidth]{PheVsCharge-4.eps}
|
---|
391 | \caption{Conversion factor $c_{phe}$ for three exemplary inner pixels (upper plots)
|
---|
392 | and three exemplary outer ones (lower plots) obtained with the extractor
|
---|
393 | {\textit{MExtractFixedWindow}} on a window size of 8 high-gain and 8 low-gain slices
|
---|
394 | (extractor \#4). }
|
---|
395 | \label{fig:linear:phevscharge4}
|
---|
396 | \end{figure}
|
---|
397 |
|
---|
398 | In this section, we test the lineary of the conversion factors FADC counts to photo-electrons:
|
---|
399 |
|
---|
400 | \begin{equation}
|
---|
401 | c_{phe} =\ <Phe> / <\widehat{S}>
|
---|
402 | \end{equation}
|
---|
403 |
|
---|
404 | As the photo-multiplier and the subsequent
|
---|
405 | optical transmission devices~\cite{david} is a linear device over a
|
---|
406 | wide dynamic range, the number of photo-electrons per charge has to remain constant over the tested
|
---|
407 | linearity region. We will show here only examples of extractors which were not already excluded in the
|
---|
408 | previous section.
|
---|
409 | \par
|
---|
410 | A first test concerns the stability of the conversion factor: mean number of averaged photo-electrons
|
---|
411 | per FADC counts over the
|
---|
412 | tested intensity region. A much more detailed investigation on the linearity will be shwon in a
|
---|
413 | separate TDAS~\cite{tdas-calibration}.
|
---|
414 |
|
---|
415 | \par
|
---|
416 | Figure~\ref{fig:linear:phevscharge4} shows the conversion factor $c_{phe}$
|
---|
417 | obtained for different light intensities
|
---|
418 | and colours for three exemplary inner and three exemplary outer pixels using a fixed window on
|
---|
419 | 8 FADC slices. Some of the pixels show a difference
|
---|
420 | between the high-gain ($<$100\ phes for the inner, $<$300\ phes for the outer pixels) and the low-gain
|
---|
421 | ($>$100\ phes for the inner, $>$300\ phes for the outer pixels) region and
|
---|
422 | a rather good stability of $c_{phe}$ for each region separately.
|
---|
423 | We conclude that the fixed window extractor \#4 is a linear extractor
|
---|
424 | for both high-gain and low-gain regions, separately.
|
---|
425 | \par
|
---|
426 |
|
---|
427 | \begin{figure}[htp]
|
---|
428 | \centering
|
---|
429 | \includegraphics[width=0.99\linewidth]{PheVsCharge-9.eps}
|
---|
430 | \caption{Conversion factor $c_{phe}$ for three exemplary inner pixels (upper plots)
|
---|
431 | and three exemplary outer ones (lower plots) obtained with the extractor
|
---|
432 | {\textit{MExtractFixedWindowSpline}}
|
---|
433 | on a window size of 8 high-gain and 8 low-gain slices (extractor \#9). }
|
---|
434 | \label{fig:linear:phevscharge9}
|
---|
435 | \end{figure}
|
---|
436 |
|
---|
437 | Figures~\ref{fig:linear:phevscharge9} and~\ref{fig:linear:phevscharge15} shows the conversion factors
|
---|
438 | using an integrated spline and a fixed window with global peak search, respectively, over
|
---|
439 | an extration window of 8 FADC slices. The same behaviour as before is obtained. These extractors are
|
---|
440 | thus linear to good approximation, for the two amplification regions, separately.
|
---|
441 | \par
|
---|
442 |
|
---|
443 | \begin{figure}[htp]
|
---|
444 | \centering
|
---|
445 | \includegraphics[width=0.99\linewidth]{PheVsCharge-15.eps}
|
---|
446 | \caption{Conversion factor $c_{phe}$ for three exemplary inner pixels (upper plots)
|
---|
447 | and three exemplary outer ones (lower plots) obtained with the extractor
|
---|
448 | {\textit{MExtractFixedWindowPeakSearch}} on a window size of 8 high-gain and 8 low-gain slices
|
---|
449 | (extractor \#15). }
|
---|
450 | \label{fig:linear:phevscharge15}
|
---|
451 | \end{figure}
|
---|
452 |
|
---|
453 | \begin{figure}[htp]
|
---|
454 | \centering
|
---|
455 | \includegraphics[width=0.99\linewidth]{PheVsCharge-20.eps}
|
---|
456 | \caption{Example of a the development of the conversion factor FADC counts to photo-electrons for three
|
---|
457 | exemplary inner pixels (upper plots) and three exemplary outer ones (lower plots) obtained with the extractor
|
---|
458 | {\textit{MExtractTimeAndChargeSlidingWindow}}
|
---|
459 | on a window size of 6 high-gain and 6 low-gain slices (extractor \#20). }
|
---|
460 | \label{fig:linear:phevscharge20}
|
---|
461 | \end{figure}
|
---|
462 |
|
---|
463 | Figure~\ref{fig:linear:phevscharge20} shows the conversion factors using a sliding window of 6 FADC slices.
|
---|
464 | The linearity is maintained like in the previous examples, except for the smallest signals where the effect
|
---|
465 | of the bias is already visible.
|
---|
466 | \par
|
---|
467 |
|
---|
468 | \begin{figure}[htp]
|
---|
469 | \centering
|
---|
470 | \includegraphics[width=0.99\linewidth]{PheVsCharge-23.eps}
|
---|
471 | \caption{Conversion factor $c_{phe}$ for three exemplary inner pixels (upper plots)
|
---|
472 | and three exemplary outer ones (lower plots) obtained with the extractor
|
---|
473 | {\textit{MExtractTimeAndChargeSpline}} with amplitude extraction (extractor \#23). }
|
---|
474 | \label{fig:linear:phevscharge23}
|
---|
475 | \end{figure}
|
---|
476 |
|
---|
477 | \begin{figure}[htp]
|
---|
478 | \centering
|
---|
479 | \includegraphics[width=0.99\linewidth]{PheVsCharge-24.eps}
|
---|
480 | \caption{Conversion factor $c_{phe}$ for three exemplary inner pixels (upper plots)
|
---|
481 | and three exemplary outer ones (lower plots) obtained with the extractor
|
---|
482 | {\textit{MExtractTimeAndChargeSpline}} with window size of 1 high-gain and 2 low-gain slices
|
---|
483 | (extractor \#24). }
|
---|
484 | \label{fig:linear:phevscharge24}
|
---|
485 | \end{figure}
|
---|
486 |
|
---|
487 | \begin{figure}[htp]
|
---|
488 | \centering
|
---|
489 | \includegraphics[width=0.99\linewidth]{PheVsCharge-25.eps}
|
---|
490 | \caption{Conversion factor $c_{phe}$ for three exemplary inner pixels (upper plots)
|
---|
491 | and three exemplary outer ones (lower plots) obtained with the extractor
|
---|
492 | {\textit{MExtractTimeAndChargeSpline}} with window size of 2 high-gain and 3 low-gain slices
|
---|
493 | (extractor \#25). }
|
---|
494 | \label{fig:linear:phevscharge25}
|
---|
495 | \end{figure}
|
---|
496 |
|
---|
497 | Figure~\ref{fig:linear:phevscharge25} shows the conversion factors using a spline
|
---|
498 | extractor with an integration window of 2 FADC slices in the high-gain and 3 FADC slices in the
|
---|
499 | low-gain. There seems to be a systematic
|
---|
500 | increase in the conversion factor in the low-gain range. In order to see if this effect is systematic,
|
---|
501 | we calculated the average of all conversion factors over the camera, separated for inner and outer
|
---|
502 | pixels (figure~\ref{fig:linear:phevschargearea25}).
|
---|
503 |
|
---|
504 |
|
---|
505 | If one uses this extractor, probably this effect will have to be corrected for.
|
---|
506 |
|
---|
507 | \par
|
---|
508 |
|
---|
509 |
|
---|
510 | \begin{figure}[htp]
|
---|
511 | \centering
|
---|
512 | \includegraphics[width=0.99\linewidth]{PheVsCharge-Area-25.eps}
|
---|
513 | \caption{Conversion factor $c_{phe}$ for three exemplary inner pixels (upper plots)
|
---|
514 | and three exemplary outer ones (lower plots) obtained with the extractor
|
---|
515 | {\textit{MExtractTimeAndChargeSpline}} with window size of 2 high-gain and 3 low-gain slices
|
---|
516 | (extractor \#25). }
|
---|
517 | \label{fig:linear:phevscharge25}
|
---|
518 | \end{figure}
|
---|
519 |
|
---|
520 | \begin{figure}[htp]
|
---|
521 | \centering
|
---|
522 | \includegraphics[width=0.99\linewidth]{PheVsCharge-30.eps}
|
---|
523 | \caption{Conversion factor $c_{phe}$ for three exemplary inner pixels (upper plots)
|
---|
524 | and three exemplary outer ones (lower plots) obtained with the extractor
|
---|
525 | {\textit{MExtractTimeAndChargeDigitalFilter}}
|
---|
526 | using a window size of 6 high-gain and 6 low-gain slices with UV-weights (extractor \#30). }
|
---|
527 | \label{fig:linear:phevscharge30}
|
---|
528 | \end{figure}
|
---|
529 |
|
---|
530 | Figure~\ref{fig:linear:phevscharge30} shows the conversion factors using a digital filter applied on 6 FADC slices with weights calculated from
|
---|
531 | the UV-calibration pulse.
|
---|
532 | One can see that all calibration blue and green calibration pulses at low and intermediate intensity fall
|
---|
533 | out of the linear region, moreover there seems to be
|
---|
534 | a systematic offset between high-gain and low-gain. These offsets have to corrected for in any way, however the loss of stability against the
|
---|
535 | exact pulse form in the high-gain is more problematic.
|
---|
536 |
|
---|
537 | \par
|
---|
538 |
|
---|
539 | \begin{figure}[htp]
|
---|
540 | \centering
|
---|
541 | \includegraphics[width=0.99\linewidth]{PheVsCharge-31.eps}
|
---|
542 | \caption{Conversion factor $c_{phe}$ for three exemplary inner pixels (upper plots)
|
---|
543 | and three exemplary outer ones (lower plots) obtained with the extractor
|
---|
544 | {\textit{MExtractTimeAndChargeDigitalFilter}} using a window size of
|
---|
545 | 4 high-gain and 4 low-gain slices (extractor \#31). }
|
---|
546 | \label{fig:linear:phevscharge31}
|
---|
547 | \end{figure}
|
---|
548 |
|
---|
549 | \clearpage
|
---|
550 |
|
---|
551 | \subsection{Time Resolution}
|
---|
552 |
|
---|
553 | The extractors \#17--32 are able to extract also the arrival time of each pulse. The calibration
|
---|
554 | delivers a fast-rising pulse, uniform over the camera in signal size and time.
|
---|
555 | We estimate the time-uniformity to better
|
---|
556 | than 300\,ps, a limit due to the different travel times of the light between inner and outer parts of the
|
---|
557 | camera. Since the calibraion does not permit a precise measurement of the absolute arrival time, we measure
|
---|
558 | the relative arrival time for every channel with respect to a reference channel (usually pixel Nr.\,1):
|
---|
559 |
|
---|
560 | \begin{equation}
|
---|
561 | \delta t_i = t_i - t_1
|
---|
562 | \end{equation}
|
---|
563 |
|
---|
564 | where $t_i$ denotes the reconstructed arrival time of pixel number $i$ and $t_1$ the reconstructed
|
---|
565 | arrival time of the reference pixel nr. 1 (software numbering). For one calibration run, one can then fill
|
---|
566 | histograms of $\delta t_i$ for each pixel and fit them to the expected Gaussian distribution. The fits
|
---|
567 | yield a mean $\mu(\delta t_i)$, comparable to
|
---|
568 | systematic offsets in the signal delay, and a sigma $\sigma(\delta t_i)$, a measure of the
|
---|
569 | combined time resolutions of pixel $i$ and pixel 1. Assuming that the PMTs and readout channels are
|
---|
570 | of a same kind, we obtain an approximate absolute time resolution of pixel $i$ by:
|
---|
571 |
|
---|
572 | \begin{equation}
|
---|
573 | t^{res}_i \approx \sigma(\delta t_i)/sqrt(2)
|
---|
574 | \end{equation}
|
---|
575 |
|
---|
576 | Figures~\ref{fig:reltimesinner10leduv} and~\ref{fig:reltimesouter10leduv} show distributions of $\delta t_i$
|
---|
577 | for
|
---|
578 | one typical inner pixel and one typical outer pixel and a non-saturating calibration pulse of UV-light,
|
---|
579 | obtained with three different extractors. One can see that the first two yield a Gaussian distribution
|
---|
580 | to a good approximation, whereas the third extractor shows a three-peak structure and cannot be fitted.
|
---|
581 | We discarded that particular extractor for this reason.
|
---|
582 |
|
---|
583 | \begin{figure}[htp]
|
---|
584 | \centering
|
---|
585 | \includegraphics[width=0.3\linewidth]{RelArrTime_Pixel97_10LedUV_Extractor32.eps}
|
---|
586 | \includegraphics[width=0.32\linewidth]{RelArrTime_Pixel97_10LedUV_Extractor23.eps}
|
---|
587 | \includegraphics[width=0.32\linewidth]{RelArrTime_Pixel97_10LedUV_Extractor17.eps}
|
---|
588 | \caption{Example of a two distributions of relative arrival times of an inner pixel with respect to
|
---|
589 | the arrival time of the reference pixel Nr. 1. The left plot shows the result using the digital filter
|
---|
590 | (extractor \#32), the central plot shows the result obtained with the half-maximum of the spline and the
|
---|
591 | right plot the result of the sliding window with a window size of 2 FADC slices (extractor \#17). A
|
---|
592 | medium sized UV-pulse (10Leds UV) has been used which does not saturate the high-gain readout channel.}
|
---|
593 | \label{fig:reltimesinner10leduv}
|
---|
594 | \end{figure}
|
---|
595 |
|
---|
596 | \begin{figure}[htp]
|
---|
597 | \centering
|
---|
598 | \includegraphics[width=0.31\linewidth]{RelArrTime_Pixel400_10LedUV_Extractor32.eps}
|
---|
599 | \includegraphics[width=0.31\linewidth]{RelArrTime_Pixel400_10LedUV_Extractor23.eps}
|
---|
600 | \includegraphics[width=0.31\linewidth]{RelArrTime_Pixel400_10LedUV_Extractor17.eps}
|
---|
601 | \caption{Example of a two distributions of relative arrival times of an outer pixel with respect to
|
---|
602 | the arrival time of the reference pixel Nr. 1. The left plot shows the result using the digital filter
|
---|
603 | (extractor \#32), the central plot shows the result obtained with the half-maximum of the spline and the
|
---|
604 | right plot the result of the sliding window with a window size of 2 FADC slices (extractor \#17). A
|
---|
605 | medium sized UV-pulse (10Leds UV) has been used which does not saturate the high-gain readout channel.}
|
---|
606 | \label{fig:reltimesouter10leduv}
|
---|
607 | \end{figure}
|
---|
608 |
|
---|
609 | Figures~\ref{fig:reltimesinner10ledsblue} and~\ref{fig:reltimesouter10ledsblue} show distributions of
|
---|
610 | $<\delta t_i>$ for
|
---|
611 | one typical inner and one typical outer pixel and a high-gain-saturating calibration pulse of blue-light,
|
---|
612 | obtained with two different extractors. One can see that the first (extractor \#23) yields a Gaussian
|
---|
613 | distribution to a good approximation.
|
---|
614 |
|
---|
615 | \begin{figure}[htp]
|
---|
616 | \centering
|
---|
617 | \includegraphics[width=0.31\linewidth]{RelArrTime_Pixel97_10LedBlue_Extractor23.eps}
|
---|
618 | \includegraphics[width=0.31\linewidth]{RelArrTime_Pixel97_10LedBlue_Extractor32.eps}
|
---|
619 | \caption{Example of a two distributions of relative arrival times of an inner pixel with respect to
|
---|
620 | the arrival time of the reference pixel Nr. 1. The left plot shows the result using the half-maximum of the spline (extractor \#23), the right plot shows the result obtained with the digital filter
|
---|
621 | (extractor \#32). A
|
---|
622 | medium sized Blue-pulse (10Leds Blue) has been used which saturates the high-gain readout channel.}
|
---|
623 | \label{fig:reltimesinner10ledsblue}
|
---|
624 | \end{figure}
|
---|
625 |
|
---|
626 |
|
---|
627 |
|
---|
628 | \begin{figure}[htp]
|
---|
629 | \centering
|
---|
630 | \includegraphics[width=0.31\linewidth]{RelArrTime_Pixel400_10LedBlue_Extractor23.eps}
|
---|
631 | \includegraphics[width=0.31\linewidth]{RelArrTime_Pixel400_10LedBlue_Extractor32.eps}
|
---|
632 | \caption{Example of a two distributions of relative arrival times of an outer pixel with respect to
|
---|
633 | the arrival time of the reference pixel Nr. 1. The left plot shows the result using the half-maximum of the spline (extractor \#23), the right plot shows the result obtained with the digital filter
|
---|
634 | (extractor \#32). A
|
---|
635 | medium sized Blue-pulse (10Leds Blue) has been used which saturates the high-gain readout channel.}
|
---|
636 | \label{fig:reltimesouter10ledsblue}
|
---|
637 | \end{figure}
|
---|
638 |
|
---|
639 | %%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
|
---|
640 |
|
---|
641 | \begin{figure}[htp]
|
---|
642 | \centering
|
---|
643 | \includegraphics[width=0.95\linewidth]{UnsuitTimeVsExtractor-5LedsUV-Colour-12.eps}
|
---|
644 | \caption{Reconstructed arrival time resolutions from a typical, not saturating calibration pulse
|
---|
645 | of colour UV, reconstructed with each of the tested arrival time extractors.
|
---|
646 | The first plots shows the time resolutions obtained for the inner pixels, the second one
|
---|
647 | for the outer pixels. Points
|
---|
648 | denote the mean of all not-excluded pixels, the error bars their RMS.}
|
---|
649 | \label{fig:time:5ledsuv}
|
---|
650 | \end{figure}
|
---|
651 |
|
---|
652 | \begin{figure}[htp]
|
---|
653 | \centering
|
---|
654 | \includegraphics[width=0.95\linewidth]{UnsuitTimeVsExtractor-1LedUV-Colour-04.eps}
|
---|
655 | \caption{Reconstructed arrival time resolutions from the lowest intensity calibration pulse
|
---|
656 | of colour UV (carrying a mean number of 4 photo-electrons),
|
---|
657 | reconstructed with each of the tested arrival time extractors.
|
---|
658 | The first plots shows the time resolutions obtained for the inner pixels, the second one
|
---|
659 | for the outer pixels. Points
|
---|
660 | denote the mean of all not-excluded pixels, the error bars their RMS.}
|
---|
661 | \label{fig:time:1leduv}
|
---|
662 | \end{figure}
|
---|
663 |
|
---|
664 | \begin{figure}[htp]
|
---|
665 | \centering
|
---|
666 | \includegraphics[width=0.95\linewidth]{UnsuitTimeVsExtractor-2LedsGreen-Colour-02.eps}
|
---|
667 | \caption{Reconstructed arrival time resolutions from a typical, not saturating calibration pulse
|
---|
668 | of colour Green, reconstructed with each of the tested arrival time extractors.
|
---|
669 | The first plots shows the time resolutions obtained for the inner pixels, the second one
|
---|
670 | for the outer pixels. Points
|
---|
671 | denote the mean of all not-excluded pixels, the error bars their RMS.}
|
---|
672 | \label{fig:time:2ledsgreen}
|
---|
673 | \end{figure}
|
---|
674 |
|
---|
675 | \begin{figure}[htp]
|
---|
676 | \centering
|
---|
677 | \includegraphics[width=0.95\linewidth]{UnsuitTimeVsExtractor-23LedsBlue-Colour-00.eps}
|
---|
678 | \caption{Reconstructed arrival time resolutions from the highest intensity calibration pulse
|
---|
679 | of colour blue, reconstructed with each of the tested arrival time extractors.
|
---|
680 | The first plots shows the time resolutions obtained for the inner pixels, the second one
|
---|
681 | for the outer pixels. Points
|
---|
682 | denote the mean of all not-excluded pixels, the error bars their RMS.}
|
---|
683 | \label{fig:time:23ledsblue}
|
---|
684 | \end{figure}
|
---|
685 |
|
---|
686 | %%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
|
---|
687 |
|
---|
688 | \begin{figure}[htp]
|
---|
689 | \centering
|
---|
690 | \includegraphics[width=0.95\linewidth]{TimeResExtractor-5LedsUV-Colour-12.eps}
|
---|
691 | \caption{Reconstructed arrival time resolutions from a typical, not saturating calibration pulse
|
---|
692 | of colour UV, reconstructed with each of the tested arrival time extractors.
|
---|
693 | The first plots shows the time resolutions obtained for the inner pixels, the second one
|
---|
694 | for the outer pixels. Points
|
---|
695 | denote the mean of all not-excluded pixels, the error bars their RMS.}
|
---|
696 | \label{fig:time:5ledsuv}
|
---|
697 | \end{figure}
|
---|
698 |
|
---|
699 | \begin{figure}[htp]
|
---|
700 | \centering
|
---|
701 | \includegraphics[width=0.95\linewidth]{TimeResExtractor-1LedUV-Colour-04.eps}
|
---|
702 | \caption{Reconstructed arrival time resolutions from the lowest intensity calibration pulse
|
---|
703 | of colour UV (carrying a mean number of 4 photo-electrons),
|
---|
704 | reconstructed with each of the tested arrival time extractors.
|
---|
705 | The first plots shows the time resolutions obtained for the inner pixels, the second one
|
---|
706 | for the outer pixels. Points
|
---|
707 | denote the mean of all not-excluded pixels, the error bars their RMS.}
|
---|
708 | \label{fig:time:1leduv}
|
---|
709 | \end{figure}
|
---|
710 |
|
---|
711 | \begin{figure}[htp]
|
---|
712 | \centering
|
---|
713 | \includegraphics[width=0.95\linewidth]{TimeResExtractor-2LedsGreen-Colour-02.eps}
|
---|
714 | \caption{Reconstructed arrival time resolutions from a typical, not saturating calibration pulse
|
---|
715 | of colour Green, reconstructed with each of the tested arrival time extractors.
|
---|
716 | The first plots shows the time resolutions obtained for the inner pixels, the second one
|
---|
717 | for the outer pixels. Points
|
---|
718 | denote the mean of all not-excluded pixels, the error bars their RMS.}
|
---|
719 | \label{fig:time:2ledsgreen}
|
---|
720 | \end{figure}
|
---|
721 |
|
---|
722 | \begin{figure}[htp]
|
---|
723 | \centering
|
---|
724 | \includegraphics[width=0.95\linewidth]{TimeResExtractor-23LedsBlue-Colour-00.eps}
|
---|
725 | \caption{Reconstructed arrival time resolutions from the highest intensity calibration pulse
|
---|
726 | of colour blue, reconstructed with each of the tested arrival time extractors.
|
---|
727 | The first plots shows the time resolutions obtained for the inner pixels, the second one
|
---|
728 | for the outer pixels. Points
|
---|
729 | denote the mean of all not-excluded pixels, the error bars their RMS.}
|
---|
730 | \label{fig:time:23ledsblue}
|
---|
731 | \end{figure}
|
---|
732 |
|
---|
733 |
|
---|
734 | \begin{figure}[htp]
|
---|
735 | \centering
|
---|
736 | \includegraphics[width=0.95\linewidth]{TimeResVsCharge-Area-21.eps}
|
---|
737 | \caption{Reconstructed mean arrival time resolutions as a function of the extracted mean number of
|
---|
738 | photo-electrons for the weighted sliding window with a window size of 8 FADC slices (extractor \#21).
|
---|
739 | Error bars denote the
|
---|
740 | spread (RMS) of the time resolutions over the investigated channels.
|
---|
741 | The marker colours show the applied
|
---|
742 | pulser colour, except for the last (green) point where all three colours were used.}
|
---|
743 | \label{fig:time:dep20}
|
---|
744 | \end{figure}
|
---|
745 |
|
---|
746 | \begin{figure}[htp]
|
---|
747 | \centering
|
---|
748 | \includegraphics[width=0.95\linewidth]{TimeResVsCharge-Area-24.eps}
|
---|
749 | \caption{Reconstructed mean arrival time resolutions as a function of the extracted mean number of
|
---|
750 | photo-electrons for the half-maximum searching spline (extractor \#23). Error bars denote the
|
---|
751 | spread (RMS) of the time resolutions over the investigated channels.
|
---|
752 | The marker colours show the applied
|
---|
753 | pulser colour, except for the last (green) point where all three colours were used.}
|
---|
754 | \label{fig:time:dep23}
|
---|
755 | \end{figure}
|
---|
756 |
|
---|
757 |
|
---|
758 | \begin{figure}[htp]
|
---|
759 | \centering
|
---|
760 | \includegraphics[width=0.95\linewidth]{TimeResVsCharge-Area-30.eps}
|
---|
761 | \caption{Reconstructed mean arrival time resolutions as a function of the extracted signal
|
---|
762 | for the digital filter with UV weights and 6 slices (extractor \#30). Error bars denote the
|
---|
763 | spread (RMS) of the time resolutions over the investigated channels.
|
---|
764 | The marker colours show the applied
|
---|
765 | pulser colour, except for the last (green) point where all three colours were used.}
|
---|
766 | \label{fig:time:dep30}
|
---|
767 | \end{figure}
|
---|
768 |
|
---|
769 |
|
---|
770 | \begin{figure}[htp]
|
---|
771 | \centering
|
---|
772 | \includegraphics[width=0.95\linewidth]{TimeResVsCharge-Area-31.eps}
|
---|
773 | \caption{Reconstructed mean arrival time resolutions as a function of the extracted signal
|
---|
774 | for the digital filter with UV weights and 4 slices (extractor \#32). Error bars denote the
|
---|
775 | spread (RMS) of the time resolutions over the investigated channels.
|
---|
776 | The marker colours show the applied
|
---|
777 | pulser colour, except for the last (green) point where all three colours were used.}
|
---|
778 | \label{fig:time:dep32}
|
---|
779 | \end{figure}
|
---|
780 |
|
---|
781 | %%% Local Variables:
|
---|
782 | %%% mode: latex
|
---|
783 | %%% TeX-master: "MAGIC_signal_reco"
|
---|
784 | %%% End:
|
---|