1 | \section{Performance \label{sec:performance}}
|
---|
2 |
|
---|
3 | \subsection{Calibration}
|
---|
4 |
|
---|
5 | In this section, we describe the tests performed using light pulses of different colour,
|
---|
6 | pulse shapes and intensities with the MAGIC LED Calibration Pulser Box \cite{hardware-manual}.
|
---|
7 | \par
|
---|
8 | The LED pulser system is able to provide fast light pulses of 3--4\,ns FWHM
|
---|
9 | with intensities ranging from 3--4 to more than 500 photo-electrons in one inner photo-multiplier of the
|
---|
10 | camera. These pulses can be produced in three colours {\textit {\bf green, blue}} and
|
---|
11 | {\textit{\bf UV}}.
|
---|
12 |
|
---|
13 | \begin{table}[htp]
|
---|
14 | \centering
|
---|
15 | \begin{tabular}{|c|c|c|c|c|c|c|}
|
---|
16 | \hline
|
---|
17 | \hline
|
---|
18 | \multicolumn{7}{|c|}{The possible pulsed light colours} \\
|
---|
19 | \hline
|
---|
20 | \hline
|
---|
21 | Colour & Wavelength & Spectral Width & Min. Nr. & Max. Nr. & Secondary & FWHM \\
|
---|
22 | & [nm] & [nm] & Phe's & Phe's & Pulses & Pulse [ns]\\
|
---|
23 | \hline
|
---|
24 | Green & 520 & 40 & 6 & 120 & yes & 3--4 \\
|
---|
25 | \hline
|
---|
26 | Blue & 460 & 30 & 6 & 500 & yes & 3--4 \\
|
---|
27 | \hline
|
---|
28 | UV & 375 & 12 & 3 & 50 & no & 2--3 \\
|
---|
29 | \hline
|
---|
30 | \hline
|
---|
31 | \end{tabular}
|
---|
32 | \caption{The pulser colours available from the calibration system}
|
---|
33 | \label{tab:pulsercolours}
|
---|
34 | \end{table}
|
---|
35 |
|
---|
36 | Table~\ref{tab:pulsercolours} lists the available colours and intensities and
|
---|
37 | figures~\ref{fig:pulseexample1leduv} and~\ref{fig:pulseexample23ledblue} show exemplary pulses
|
---|
38 | as registered by the FADCs.
|
---|
39 | Whereas the UV-pulse is very stable, the green and blue pulses show sometimes smaller secondary
|
---|
40 | pulses after about 10--40\,ns from the main pulse.
|
---|
41 | One can see that the very stable UV-pulses are unfortunately only available in such intensities as to
|
---|
42 | not saturate the high-gain readout channel. However, the brightest combination of light pulses easily
|
---|
43 | saturates all channels in the camera, but does not reach a saturation of the low-gain readout.
|
---|
44 | \par
|
---|
45 | Our tests can be classified into three subsections:
|
---|
46 |
|
---|
47 | \begin{enumerate}
|
---|
48 | \item Un-calibrated pixels and events: These tests measure the percentage of failures of the extractor
|
---|
49 | resulting either in a pixel declared as un-calibrated or in an event which produces a signal ouside
|
---|
50 | of the expected Gaussian distribution.
|
---|
51 | \item Number of photo-electrons: These tests measure the reconstructed numbers of photo-electrons, their
|
---|
52 | spread over the camera and the ratio of the obtained mean values for outer and inner pixels, respectively.
|
---|
53 | \item Linearity tests: These tests measure the linearity of the extractor with respect to pulses of
|
---|
54 | different intensity and colour.
|
---|
55 | \item Time resolution: These tests show the time resolution and stability obtained with different
|
---|
56 | intensities and colours.
|
---|
57 | \end{enumerate}
|
---|
58 |
|
---|
59 | \begin{figure}[htp]
|
---|
60 | \centering
|
---|
61 | \includegraphics[width=0.48\linewidth]{1LedUV_Pulse_Inner.eps}
|
---|
62 | \includegraphics[width=0.48\linewidth]{1LedUV_Pulse_Outer.eps}
|
---|
63 | \caption{Example of a calibration pulse from the lowest available intensity (1\,Led UV).
|
---|
64 | The left plot shows the signal obtained in an inner pixel, the right one the signal in an outer pixel.
|
---|
65 | Note that the pulse height fluctuates much more than suggested from these pictures. Especially, a
|
---|
66 | zero-pulse is also possible.}
|
---|
67 | \label{fig:pulseexample1leduv}
|
---|
68 | \end{figure}
|
---|
69 |
|
---|
70 | \begin{figure}[htp]
|
---|
71 | \centering
|
---|
72 | \includegraphics[width=0.48\linewidth]{23LedsBlue_Pulse_Inner.eps}
|
---|
73 | \includegraphics[width=0.48\linewidth]{23LedsBlue_Pulse_Outer.eps}
|
---|
74 | \caption{Example of a calibration pulse from the highest available mono-chromatic intensity (23\,Leds Blue).
|
---|
75 | The left plot shows the signal obtained in an inner pixel, the right one the signal in an outer pixel.
|
---|
76 | One the left side of both plots, the (saturated) high-gain channel is visible,
|
---|
77 | on the right side from FADC slice 18 on,
|
---|
78 | the delayed low-gain
|
---|
79 | pulse appears. Note that in the left plot, there is a secondary pulses visible in the tail of the
|
---|
80 | high-gain pulse. }
|
---|
81 | \label{fig:pulseexample23ledblue}
|
---|
82 | \end{figure}
|
---|
83 |
|
---|
84 | We used data taken on the 7$^{th}$ of June, 2004 with different pulser LED combinations, each taken with
|
---|
85 | 16384 events. The corresponding MAGIC data run numbers range from nr. 31741 to 31772. These data was taken
|
---|
86 | before the latest camera repair access which resulted in a replacement of about 2\% of the pixels known to be
|
---|
87 | mal-functionning at that time.
|
---|
88 | There is thus a lower limit to the number of un-calibrated pixels of about 1.5--2\% of known
|
---|
89 | mal-functionning photo-multipliers.
|
---|
90 | \par
|
---|
91 | Although we had looked at and tested all colour and extractor combinations resulting from these data,
|
---|
92 | we refrain ourselves to show here only exemplary behaviour and results of extractors.
|
---|
93 | All plots, including those which are not displayed in this TDAS, can be retrieved from the following
|
---|
94 | locations:
|
---|
95 |
|
---|
96 | \begin{verbatim}
|
---|
97 | http://www.magic.ifae.es/~markus/pheplots/
|
---|
98 | http://www.magic.ifae.es/~markus/timeplots/
|
---|
99 | \end{verbatim}
|
---|
100 |
|
---|
101 | %%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
|
---|
102 |
|
---|
103 | \subsubsection{Un-Calibrated Pixels and Events}
|
---|
104 |
|
---|
105 | The MAGIC calibration software incorporates a series of checks to sort out mal-functionning pixels.
|
---|
106 | Except for the software bug searching criteria, the following exclusion criteria can apply:
|
---|
107 |
|
---|
108 | \begin{enumerate}
|
---|
109 | \item The reconstructed mean signal is less than 2.5 times the extractor resolution $R$ from zero.
|
---|
110 | (2.5 Pedestal RMS in the case of the simple fixed window extractors, see section~\ref{sec:pedestals}).
|
---|
111 | This criterium essentially cuts out
|
---|
112 | dead pixels.
|
---|
113 | \item The reconstructed mean signal error is smaller than its value. This criterium cuts out
|
---|
114 | signal distributions which fluctuate so much that their RMS is bigger than its mean value. This
|
---|
115 | criterium cuts out ``ringing'' pixels or mal-functionning extractors.
|
---|
116 | \item The reconstructed mean number of photo-electrons lies 4.5 sigma outside
|
---|
117 | the distribution of photo-electrons obtained with the inner or outer pixels in the camera, respectively.
|
---|
118 | This criterium cuts out pixels channels with apparently deviating (hardware) behaviour compared to
|
---|
119 | the rest of the camera readout\footnote{This criteria is not applied any more in the standard analysis,
|
---|
120 | although here, we kept using it}.
|
---|
121 | \item All pixels with reconstructed negative mean signal or with a
|
---|
122 | mean numbers of photo-electrons smaller than one. Pixels with a negative pedestal RMS subtracted
|
---|
123 | sigma occur, especially when stars are focussed onto that pixel during the pedestal taking (resulting
|
---|
124 | in a large pedestal RMS), but have moved to another pixel during the calibration run. In this case, the
|
---|
125 | number of photo-electrons would result artificially negative. If these
|
---|
126 | channels do not show any other deviating behaviour, their number of photo-electrons gets replaced by the
|
---|
127 | mean number of photo-electrons in the camera, and the channel is further calibrated as normal.
|
---|
128 | \end{enumerate}
|
---|
129 |
|
---|
130 | Moreover, the number of events are counted which have been reconstructed outside a 5 sigma region
|
---|
131 | from the mean signal. These events are called ``outliers''. Figure~\ref{fig:outlier} shows a typical
|
---|
132 | outlier obtained with the digital filter applied to a low-gain signal.
|
---|
133 |
|
---|
134 | \begin{figure}[htp]
|
---|
135 | \centering
|
---|
136 | \includegraphics[width=0.95\linewidth]{Outlier.eps}
|
---|
137 | \caption{Example of an event classified as ``un-calibrated''. The histogram has been obtained
|
---|
138 | using the digital filter (extractor \#32) applied to a high-intensity blue pulse (run 31772).
|
---|
139 | The event marked as ``outlier'' clearly has been mis-reconstructed. It lies outside the 5 sigma
|
---|
140 | region from the fitted mean.}
|
---|
141 | \label{fig:outlier}
|
---|
142 | \end{figure}
|
---|
143 |
|
---|
144 | The following figures~\ref{fig:unsuited:5ledsuv},~\ref{fig:unsuited:1leduv},~\ref{fig:unsuited:2ledsgreen}
|
---|
145 | and~\ref{fig:unsuited:23ledsblue} show the resulting numbers of un-calibrated pixels and events for
|
---|
146 | different colours and intensities. Because there is a strong anti-correlation between the number of
|
---|
147 | excluded channels and the number of outliers per event, we have chosen to show these numbers together.
|
---|
148 |
|
---|
149 | \par
|
---|
150 |
|
---|
151 | \begin{figure}[htp]
|
---|
152 | \centering
|
---|
153 | \includegraphics[height=0.95\textheight]{UnsuitVsExtractor-5LedsUV-Colour-13.eps}
|
---|
154 | \caption{Uncalibrated pixels and pixels outside of the Gaussian distribution for a typical calibration
|
---|
155 | pulse of UV-light which does not saturate the high-gain readout.}
|
---|
156 | \label{fig:unsuited:5ledsuv}
|
---|
157 | \end{figure}
|
---|
158 |
|
---|
159 | \begin{figure}[htp]
|
---|
160 | \centering
|
---|
161 | \includegraphics[height=0.95\textheight]{UnsuitVsExtractor-1LedUV-Colour-04.eps}
|
---|
162 | \caption{Uncalibrated pixels and pixels outside of the Gaussian distribution for a very low
|
---|
163 | intensity pulse.}
|
---|
164 | \label{fig:unsuited:1leduv}
|
---|
165 | \end{figure}
|
---|
166 |
|
---|
167 | \begin{figure}[htp]
|
---|
168 | \centering
|
---|
169 | \includegraphics[height=0.95\textheight]{UnsuitVsExtractor-2LedsGreen-Colour-02.eps}
|
---|
170 | \caption{Uncalibrated pixels and pixels outside of the Gaussian distribution for a typical green pulse.}
|
---|
171 | \label{fig:unsuited:2ledsgreen}
|
---|
172 | \end{figure}
|
---|
173 |
|
---|
174 | \begin{figure}[htp]
|
---|
175 | \centering
|
---|
176 | \includegraphics[height=0.95\textheight]{UnsuitVsExtractor-23LedsBlue-Colour-00.eps}
|
---|
177 | \caption{Uncalibrated pixels and pixels outside of the Gaussian distribution for a high-intensity blue pulse.}
|
---|
178 | \label{fig:unsuited:23ledsblue}
|
---|
179 | \end{figure}
|
---|
180 |
|
---|
181 | One can see that in general, big extraction windows raise the
|
---|
182 | number of un-calibrated pixels and are thus less stable. Especially for the very low-intensity
|
---|
183 | \textit{\bf 1Led\,UV}-pulse, the big extraction windows summing 8 or more slices, cannot calibrate more
|
---|
184 | than 50\%
|
---|
185 | of the inner pixels (fig.~\ref{fig:unsuited:1leduv}). This is an expected behavior since big windows
|
---|
186 | add up more noise which in turn makes the search for the small signal more difficult.
|
---|
187 | \par
|
---|
188 | \ldots {\bf WHICH EXTRACTOR HAS THE LEAST NUMBER OF EXCLUDED PIXELS ???}
|
---|
189 | \par
|
---|
190 | In general, one can also find that all ``sliding window''-algorithms (extractors \#17-32) discard
|
---|
191 | less pixels than the ``fixed window''-ones (extractors \#1--16). The digital filter with
|
---|
192 | the correct weights (extractors \#30-33) discards the least number of pixels and is also robust against
|
---|
193 | slight modifications of its weights (extractors \#28--30). The robustness gets lost when the high-gain and
|
---|
194 | low-gain weights are inverted (extractors \#31--39, see fig.~\ref{fig:unsuited:23ledsblue}).
|
---|
195 | \par
|
---|
196 | Also the ``spline'' algorithms on small
|
---|
197 | windows (extractors \#23--25) discard less pixels than the previous extractors.
|
---|
198 | \par
|
---|
199 | It seems also that the spline algorithm extracting the amplitude of the signal produces an over-proportional
|
---|
200 | number of excluded events in the low-gain. The same, however in a less significant manner, holds for
|
---|
201 | the digital filter with high-low-gain inverted weights. The limit of stability with respect to
|
---|
202 | changes in the pulse form seems to be reached, there.
|
---|
203 | \par
|
---|
204 | Concerning the numbers of outliers, one can conclude that in general, the numbers are very low never exceeding
|
---|
205 | 0.1\% except for the ampltiude-extracting spline which seems to mis-reconstruct a certain type of events.
|
---|
206 | \par
|
---|
207 | In conclusion, already this first test excludes all extractors with too big window sizes because
|
---|
208 | they are not able to extract cleanly small signals produced by about 4 photo-electrons. Moreover,
|
---|
209 | some extractors do not reproduce the signals as expected in the low-gain.
|
---|
210 | The excluded extractors are:
|
---|
211 | \begin{itemize}
|
---|
212 | \item: MExtractFixedWindow Nr. 3--5
|
---|
213 | \item: MExtractFixedWindowSpline Nr. 6--11 (all)
|
---|
214 | \item: MExtractFixedWindowPeakSearch Nr. 14--16
|
---|
215 | \item: MExtractTimeAndChargeSlidingWindow Nr. 21--22
|
---|
216 | \item: MExtractTimeAndChargeSpline Nr. 23 and 27
|
---|
217 | \end{itemize}
|
---|
218 |
|
---|
219 | %%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
|
---|
220 |
|
---|
221 | \subsubsection{Number of Photo-Electrons \label{sec:photo-electrons}}
|
---|
222 |
|
---|
223 | Assuming that the readout chain adds only negligible noise to the one
|
---|
224 | introduced by the photo-multiplier itself, one can make the assumption that the variance of the
|
---|
225 | true (non-extracted) signal $ST$ is the amplified Poisson variance of the number of photo-electrons,
|
---|
226 | multiplied with the excess noise of the photo-multiplier which itself is
|
---|
227 | characterized by the excess-noise factor $F$.
|
---|
228 |
|
---|
229 | \begin{equation}
|
---|
230 | Var(ST) = F^2 \cdot Var(N_{phe}) \cdot \frac{<ST>^2}{<N_{phe}>^2}
|
---|
231 | \label{eq:excessnoise}
|
---|
232 | \end{equation}
|
---|
233 |
|
---|
234 | After introducing the effect of the night-sky background (eq.~\ref{eq:rmssubtraction})
|
---|
235 | in formula~\ref{eq:excessnoise} and assuming that the variance of the number of photo-electrons is equal
|
---|
236 | to the mean number of photo-electrons (because of the Poisson distribution),
|
---|
237 | one obtains an expression to retrieve the mean number of photo-electrons impinging on the pixel from the
|
---|
238 | mean extracted signal $<SE>$, its variance $Var(SE)$ and the RMS of the extracted signal obtained from
|
---|
239 | pure pedestal runs $R$ (see section~\ref{sec:determiner}):
|
---|
240 |
|
---|
241 | \begin{equation}
|
---|
242 | <N_{phe}> \approx F^2 \cdot \frac{<SE>^2}{Var(SE) - R^2}
|
---|
243 | \label{eq:pheffactor}
|
---|
244 | \end{equation}
|
---|
245 |
|
---|
246 | In theory, eq.~\ref{eq:pheffactor} must not depend on the extractor! Effectively, we will use it to test the
|
---|
247 | quality of our extractors by requiring that a valid extractor yields the same number of photo-electrons
|
---|
248 | for all pixels of a same type and does not deviate from the number obtained with other extractors.
|
---|
249 | As the camera is flat-fielded, but the number of photo-electrons impinging on an inner and an outer pixel is
|
---|
250 | different, we also use the ratio of the mean numbers of photo-electrons from the outer pixels to the one
|
---|
251 | obtained from the inner pixels as a test variable. In the ideal case, it should always yield its central
|
---|
252 | value of about 2.6$\pm$0.1~\cite{michele-diploma}.
|
---|
253 | \par
|
---|
254 | In our case, there is an additional complication due to the fact that the green and blue coloured light pulses
|
---|
255 | show secondary pulses which destroy the Poisson behaviour of the number of photo-electrons. We will
|
---|
256 | have to split our sample of extractors into those being affected by the secondary pulses and those
|
---|
257 | being immune to this effect.
|
---|
258 | \par
|
---|
259 | Figures~\ref{fig:phe:5ledsuv},~\ref{fig:phe:1leduv},~\ref{fig:phe:2ledsgreen}~and~\ref{fig:phe:23ledsblue} show
|
---|
260 | some of the obtained results. Although one can see a rather good stability for the standard
|
---|
261 | {\textit{\bf 5\,Leds\,UV}}\ pulse, except for the extractors {\textit{\bf MExtractFixedWindowPeakSearch}}, initialized
|
---|
262 | with an extraction window of 2 slices and {\textit{\bf MExtractTimeAndChargeDigitalFilter}}, initialized with
|
---|
263 | an extraction window of 4 slices (extractor \#29).
|
---|
264 | \par
|
---|
265 | There is a considerable difference for all shown non-standard pulses. Especially the pulses from green
|
---|
266 | and blue LEDs
|
---|
267 | show a clear dependency of the number of photo-electrons on the extraction window. Only the largest
|
---|
268 | extraction windows seem to catch the entire range of (jittering) secondary pulses and get the ratio
|
---|
269 | of outer vs. inner pixels right. However, they (obviously) over-estimate the number of photo-electrons
|
---|
270 | in the primary pulse.
|
---|
271 | \par
|
---|
272 | The strongest discrepancy is observed in the low-gain extraction (fig.~\ref{fig:phe:23ledsblue}) where all
|
---|
273 | fixed window extractors with too small extraction windows fail to reconstruct the correct numbers.
|
---|
274 | This has to do with the fact that
|
---|
275 | the fixed window extractors fail to do catch a significant part of the (larger) pulse because of the
|
---|
276 | 1~FADC slice event-to-event jitter.
|
---|
277 |
|
---|
278 |
|
---|
279 | \begin{figure}[htp]
|
---|
280 | \centering
|
---|
281 | \includegraphics[height=0.92\textheight]{PheVsExtractor-5LedsUV-Colour-13.eps}
|
---|
282 | \caption{Number of photo-electrons from a typical, not saturating calibration pulse of colour UV,
|
---|
283 | reconstructed with each of the tested signal extractors.
|
---|
284 | The first plots shows the number of photo-electrons obtained for the inner pixels, the second one
|
---|
285 | for the outer pixels and the third shows the ratio of the mean number of photo-electrons for the
|
---|
286 | outer pixels divided by the mean number of photo-electrons for the inner pixels. Points
|
---|
287 | denote the mean of all not-excluded pixels, the error bars their RMS.}
|
---|
288 | \label{fig:phe:5ledsuv}
|
---|
289 | \end{figure}
|
---|
290 |
|
---|
291 | \begin{figure}[htp]
|
---|
292 | \centering
|
---|
293 | \includegraphics[height=0.92\textheight]{PheVsExtractor-1LedUV-Colour-04.eps}
|
---|
294 | \caption{Number of photo-electrons from a typical, very low-intensity calibration pulse of colour UV,
|
---|
295 | reconstructed with each of the tested signal extractors.
|
---|
296 | The first plots shows the number of photo-electrons obtained for the inner pixels, the second one
|
---|
297 | for the outer pixels and the third shows the ratio of the mean number of photo-electrons for the
|
---|
298 | outer pixels divided by the mean number of photo-electrons for the inner pixels. Points
|
---|
299 | denote the mean of all not-excluded pixels, the error bars their RMS.}
|
---|
300 | \label{fig:phe:1leduv}
|
---|
301 | \end{figure}
|
---|
302 |
|
---|
303 | \begin{figure}[htp]
|
---|
304 | \centering
|
---|
305 | \includegraphics[height=0.92\textheight]{PheVsExtractor-2LedsGreen-Colour-02.eps}
|
---|
306 | \caption{Number of photo-electrons from a typical, not saturating calibration pulse of colour green,
|
---|
307 | reconstructed with each of the tested signal extractors.
|
---|
308 | The first plots shows the number of photo-electrons obtained for the inner pixels, the second one
|
---|
309 | for the outer pixels and the third shows the ratio of the mean number of photo-electrons for the
|
---|
310 | outer pixels divided by the mean number of photo-electrons for the inner pixels. Points
|
---|
311 | denote the mean of all not-excluded pixels, the error bars their RMS.}
|
---|
312 | \label{fig:phe:2ledsgreen}
|
---|
313 | \end{figure}
|
---|
314 |
|
---|
315 |
|
---|
316 | \begin{figure}[htp]
|
---|
317 | \centering
|
---|
318 | \includegraphics[height=0.92\textheight]{PheVsExtractor-23LedsBlue-Colour-00.eps}
|
---|
319 | \caption{Number of photo-electrons from a typical, high-gain saturating calibration pulse of colour blue,
|
---|
320 | reconstructed with each of the tested signal extractors.
|
---|
321 | The first plots shows the number of photo-electrons obtained for the inner pixels, the second one
|
---|
322 | for the outer pixels and the third shows the ratio of the mean number of photo-electrons for the
|
---|
323 | outer pixels divided by the mean number of photo-electrons for the inner pixels. Points
|
---|
324 | denote the mean of all not-excluded pixels, the error bars their RMS.}
|
---|
325 | \label{fig:phe:23ledsblue}
|
---|
326 | \end{figure}
|
---|
327 |
|
---|
328 | One can see that all extractors using a large window belong to the class of extractors being affected
|
---|
329 | by the secondary pulses, except for the digital filter. The only exception to this rule is the digital filter
|
---|
330 | which - despite of its 6 slices extraction window - seems to filter out all the secondary pulses.
|
---|
331 | \par
|
---|
332 | The extractor {\textit{\bf MExtractFixedWindowPeakSearch}} at low extraction windows apparently yields chronically low
|
---|
333 | numbers of photo-electrons. This is due to the fact that the decision to fix the extraction window is
|
---|
334 | made sometimes by an inner pixel and sometimes by an outer one since the camera is flat-fielded and the
|
---|
335 | pixel carrying the largest non-saturated peak-search window is more or less found by a random signal
|
---|
336 | fluctuation. However, inner and outer pixels have a systematic offset of about 0.5 to 1 FADC slices.
|
---|
337 | Thus, the extraction fluctuates artificially for one given channel which results in a systematically
|
---|
338 | large variance and thus in a systematically low reconstructed number of photo-electrons. This test thus
|
---|
339 | excludes the extractors \#11--13.
|
---|
340 | \par
|
---|
341 | Moreover, one can see that the extractors applying a small fixed window do not get the ratio of
|
---|
342 | photo-electrons correctly between outer to inner pixels for the green and blue pulses.
|
---|
343 | \par
|
---|
344 | The extractor {\textit{\bf MExtractTimeAndChargeDigitalFilter}} seems to be stable against modifications in the
|
---|
345 | exact form of the weights in the high-gain readout channel since all applied weights yield about
|
---|
346 | the same number of photo-electrons and the same ratio of outer vs. inner pixels. This statement does not
|
---|
347 | hold any more for the low-gain, as can be seen in figure~\ref{fig:phe:23ledsblue}. There, the application
|
---|
348 | of high-gain weights to the low-gain signal (extractors \#34--39) produces a too low number of photo-electrons
|
---|
349 | and also a too low ratio of outer vs. inner pixels.
|
---|
350 | \par
|
---|
351 | All sliding window and spline algorithms yield a stable ratio of outer vs. inner pixels in the low-gain,
|
---|
352 | however the effect of raising the number of photo-electrons with the extraction window is very pronounced.
|
---|
353 | Note that in figure~\ref{fig:phe:23ledsblue}, the number of photo-electrons rises by about a factor 1.4,
|
---|
354 | which is slightly higher than in the case of the high-gain channel (figure~\ref{fig:phe:2ledsgreen}).
|
---|
355 | \par
|
---|
356 | Concluding, there is no fixed window extractor yielding the correct number of photo-electrons
|
---|
357 | for the low-gain, except for the largest extraction window of 8 and 10 low-gain slices.
|
---|
358 | Either the number of photo-electrons itself is wrong or the ratio of outer vs. inner pixels is
|
---|
359 | not correct. All sliding window algorithms seem to reproduce the correct numbers if one takes into
|
---|
360 | account the after-pulse behaviour of the light pulser itself. The digital filter seems to be
|
---|
361 | unstable against exchanging the pulse form to match the slimmer high-gain pulses, though.
|
---|
362 |
|
---|
363 | \par
|
---|
364 | \ldots {\textit{\bf EXCLUDED : CW4, UV4 No stability High-gain vs. LoGain}}
|
---|
365 | \par
|
---|
366 |
|
---|
367 | \subsubsection{Linearity Tests}
|
---|
368 |
|
---|
369 | In this section, we test the lineary of the extractors. As the photo-multiplier and the subsequent
|
---|
370 | optical transmission devices~\cite{david} is a linear device over a
|
---|
371 | wide dynamic range, the number of photo-electrons per charge has to remain constant over the tested
|
---|
372 | linearity region. We will show here only examples of extractors which were not already excluded in the
|
---|
373 | previous section.
|
---|
374 | \par
|
---|
375 | A first test concerns the stability of the conversion factor: mean number of averaged photo-electrons
|
---|
376 | per FADC counts over the
|
---|
377 | tested intensity region. A much more detailed investigation on the linearity will be shwon in a
|
---|
378 | separate TDAS~\cite{tdas-calibration}.
|
---|
379 |
|
---|
380 |
|
---|
381 | \begin{figure}[htp]
|
---|
382 | \centering
|
---|
383 | \includegraphics[width=0.95\linewidth]{PheVsCharge-3.eps}
|
---|
384 | \caption{Example of a the development of the conversion factor FADC counts to photo-electrons for two
|
---|
385 | exemplary inner pixels (upper plots) and two exemplary outer ones (lower plots).
|
---|
386 | A fixed window extractor on a window size of 6 high-gain and 6 low-gain slices has been used (extractor \#3). }
|
---|
387 | \label{fig:linear:phevscharge3}
|
---|
388 | \end{figure}
|
---|
389 |
|
---|
390 | \begin{figure}[htp]
|
---|
391 | \centering
|
---|
392 | \includegraphics[width=0.95\linewidth]{PheVsCharge-8.eps}
|
---|
393 | \caption{Example of a the development of the conversion factor FADC counts to photo-electrons for two
|
---|
394 | exemplary inner pixels (upper plots) and two exemplary outer ones (lower plots).
|
---|
395 | A fixed window spline extractor on a window size of 6 high-gain and 6 low-gain slices has been used
|
---|
396 | (extractor \#8). }
|
---|
397 | \label{fig:linear:phevscharge8}
|
---|
398 | \end{figure}
|
---|
399 |
|
---|
400 | \begin{figure}[htp]
|
---|
401 | \centering
|
---|
402 | \includegraphics[width=0.95\linewidth]{PheVsCharge-14.eps}
|
---|
403 | \caption{Example of a the development of the conversion factor FADC counts to photo-electrons for two
|
---|
404 | exemplary inner pixels (upper plots) and two exemplary outer ones (lower plots).
|
---|
405 | A fixed window peak search extractor on a window size of 6 high-gain and 6 low-gain slices has been used
|
---|
406 | (extractor \#14). }
|
---|
407 | \label{fig:linear:phevscharge14}
|
---|
408 | \end{figure}
|
---|
409 |
|
---|
410 | \begin{figure}[htp]
|
---|
411 | \centering
|
---|
412 | \includegraphics[width=0.95\linewidth]{PheVsCharge-20.eps}
|
---|
413 | \caption{Example of a the development of the conversion factor FADC counts to photo-electrons for two
|
---|
414 | exemplary inner pixels (upper plots) and two exemplary outer ones (lower plots).
|
---|
415 | A sliding window extractor on a window size of 6 high-gain and 6 low-gain slices has been used
|
---|
416 | (extractor \#20). }
|
---|
417 | \label{fig:linear:phevscharge20}
|
---|
418 | \end{figure}
|
---|
419 |
|
---|
420 | \begin{figure}[htp]
|
---|
421 | \centering
|
---|
422 | \includegraphics[width=0.95\linewidth]{PheVsCharge-25.eps}
|
---|
423 | \caption{Example of a the development of the conversion factor FADC counts to photo-electrons for two
|
---|
424 | exemplary inner pixels (upper plots) and two exemplary outer ones (lower plots).
|
---|
425 | An integrating spline extractor on a sliding window and a window size of 2 high-gain and 3 low-gain slices
|
---|
426 | has been used (extractor \#25). }
|
---|
427 | \label{fig:linear:phevscharge25}
|
---|
428 | \end{figure}
|
---|
429 |
|
---|
430 | \begin{figure}[htp]
|
---|
431 | \centering
|
---|
432 | \includegraphics[width=0.95\linewidth]{PheVsCharge-27.eps}
|
---|
433 | \caption{Example of a the development of the conversion factor FADC counts to photo-electrons for two
|
---|
434 | exemplary inner pixels (upper plots) and two exemplary outer ones (lower plots).
|
---|
435 | An integrating spline extractor on a sliding window and a window size of 6 high-gain and 7 low-gain slices
|
---|
436 | has been used (extractor \#27). }
|
---|
437 | \label{fig:linear:phevscharge27}
|
---|
438 | \end{figure}
|
---|
439 |
|
---|
440 | \begin{figure}[htp]
|
---|
441 | \centering
|
---|
442 | \includegraphics[width=0.95\linewidth]{PheVsCharge-30.eps}
|
---|
443 | \caption{Example of a the development of the conversion factor FADC counts to photo-electrons for two
|
---|
444 | exemplary inner pixels (upper plots) and two exemplary outer ones (lower plots).
|
---|
445 | A digital filter extractor on a window size of 6 high-gain and 6 low-gain slices has been used
|
---|
446 | with UV-weights (extractor \#30). }
|
---|
447 | \label{fig:linear:phevscharge30}
|
---|
448 | \end{figure}
|
---|
449 |
|
---|
450 | \begin{figure}[htp]
|
---|
451 | \centering
|
---|
452 | \includegraphics[width=0.95\linewidth]{PheVsCharge-31.eps}
|
---|
453 | \caption{Example of a the development of the conversion factor FADC counts to photo-electrons for two
|
---|
454 | exemplary inner pixels (upper plots) and two exemplary outer ones (lower plots).
|
---|
455 | A digital filter extractor on a window size of 4 high-gain and 4 low-gain slices has been used
|
---|
456 | (extractor \#31). }
|
---|
457 | \label{fig:linear:phevscharge31}
|
---|
458 | \end{figure}
|
---|
459 |
|
---|
460 |
|
---|
461 |
|
---|
462 | \subsubsection{Time Resolution}
|
---|
463 |
|
---|
464 | The extractors \#17--32 are able to extract also the arrival time of each pulse. The calibration
|
---|
465 | delivers a fast-rising pulse, uniform over the camera in signal size and time.
|
---|
466 | We estimate the time-uniformity to better
|
---|
467 | than 300\,ps, a limit due to the different travel times of the light between inner and outer parts of the
|
---|
468 | camera. Since the calibraion does not permit a precise measurement of the absolute arrival time, we measure
|
---|
469 | the relative arrival time for every channel with respect to a reference channel (usually pixel Nr.\,1):
|
---|
470 |
|
---|
471 | \begin{equation}
|
---|
472 | \delta t_i = t_i - t_1
|
---|
473 | \end{equation}
|
---|
474 |
|
---|
475 | where $t_i$ denotes the reconstructed arrival time of pixel number $i$ and $t_1$ the reconstructed
|
---|
476 | arrival time of the reference pixel nr. 1 (software numbering). For one calibration run, one can then fill
|
---|
477 | histograms of $\delta t_i$ for each pixel and fit them to the expected Gaussian distribution. The fits
|
---|
478 | yield a mean $\mu(\delta t_i)$, comparable to
|
---|
479 | systematic offsets in the signal delay, and a sigma $\sigma(\delta t_i)$, a measure of the
|
---|
480 | combined time resolutions of pixel $i$ and pixel 1. Assuming that the PMTs and readout channels are
|
---|
481 | of a same kind, we obtain an approximate absolute time resolution of pixel $i$ by:
|
---|
482 |
|
---|
483 | \begin{equation}
|
---|
484 | t^{res}_i \approx \sigma(\delta t_i)/sqrt(2)
|
---|
485 | \end{equation}
|
---|
486 |
|
---|
487 | Figures~\ref{fig:reltimesinner10leduv} and~\ref{fig:reltimesouter10leduv} show distributions of $\delta t_i$
|
---|
488 | for
|
---|
489 | one typical inner pixel and one typical outer pixel and a non-saturating calibration pulse of UV-light,
|
---|
490 | obtained with three different extractors. One can see that the first two yield a Gaussian distribution
|
---|
491 | to a good approximation, whereas the third extractor shows a three-peak structure and cannot be fitted.
|
---|
492 | We discarded that particular extractor for this reason.
|
---|
493 |
|
---|
494 | \begin{figure}[htp]
|
---|
495 | \centering
|
---|
496 | \includegraphics[width=0.3\linewidth]{RelArrTime_Pixel97_10LedUV_Extractor32.eps}
|
---|
497 | \includegraphics[width=0.32\linewidth]{RelArrTime_Pixel97_10LedUV_Extractor23.eps}
|
---|
498 | \includegraphics[width=0.32\linewidth]{RelArrTime_Pixel97_10LedUV_Extractor17.eps}
|
---|
499 | \caption{Example of a two distributions of relative arrival times of an inner pixel with respect to
|
---|
500 | the arrival time of the reference pixel Nr. 1. The left plot shows the result using the digital filter
|
---|
501 | (extractor \#32), the central plot shows the result obtained with the half-maximum of the spline and the
|
---|
502 | right plot the result of the sliding window with a window size of 2 FADC slices (extractor \#17). A
|
---|
503 | medium sized UV-pulse (10Leds UV) has been used which does not saturate the high-gain readout channel.}
|
---|
504 | \label{fig:reltimesinner10leduv}
|
---|
505 | \end{figure}
|
---|
506 |
|
---|
507 | \begin{figure}[htp]
|
---|
508 | \centering
|
---|
509 | \includegraphics[width=0.31\linewidth]{RelArrTime_Pixel400_10LedUV_Extractor32.eps}
|
---|
510 | \includegraphics[width=0.31\linewidth]{RelArrTime_Pixel400_10LedUV_Extractor23.eps}
|
---|
511 | \includegraphics[width=0.31\linewidth]{RelArrTime_Pixel400_10LedUV_Extractor17.eps}
|
---|
512 | \caption{Example of a two distributions of relative arrival times of an outer pixel with respect to
|
---|
513 | the arrival time of the reference pixel Nr. 1. The left plot shows the result using the digital filter
|
---|
514 | (extractor \#32), the central plot shows the result obtained with the half-maximum of the spline and the
|
---|
515 | right plot the result of the sliding window with a window size of 2 FADC slices (extractor \#17). A
|
---|
516 | medium sized UV-pulse (10Leds UV) has been used which does not saturate the high-gain readout channel.}
|
---|
517 | \label{fig:reltimesouter10leduv}
|
---|
518 | \end{figure}
|
---|
519 |
|
---|
520 | Figures~\ref{fig:reltimesinner10ledsblue} and~\ref{fig:reltimesouter10ledsblue} show distributions of
|
---|
521 | $<\delta t_i>$ for
|
---|
522 | one typical inner and one typical outer pixel and a high-gain-saturating calibration pulse of blue-light,
|
---|
523 | obtained with two different extractors. One can see that the first (extractor \#23) yields a Gaussian
|
---|
524 | distribution to a good approximation.
|
---|
525 |
|
---|
526 | \begin{figure}[htp]
|
---|
527 | \centering
|
---|
528 | \includegraphics[width=0.31\linewidth]{RelArrTime_Pixel97_10LedBlue_Extractor23.eps}
|
---|
529 | \includegraphics[width=0.31\linewidth]{RelArrTime_Pixel97_10LedBlue_Extractor32.eps}
|
---|
530 | \caption{Example of a two distributions of relative arrival times of an inner pixel with respect to
|
---|
531 | the arrival time of the reference pixel Nr. 1. The left plot shows the result using the half-maximum of the spline (extractor \#23), the right plot shows the result obtained with the digital filter
|
---|
532 | (extractor \#32). A
|
---|
533 | medium sized Blue-pulse (10Leds Blue) has been used which saturates the high-gain readout channel.}
|
---|
534 | \label{fig:reltimesinner10ledsblue}
|
---|
535 | \end{figure}
|
---|
536 |
|
---|
537 |
|
---|
538 |
|
---|
539 | \begin{figure}[htp]
|
---|
540 | \centering
|
---|
541 | \includegraphics[width=0.31\linewidth]{RelArrTime_Pixel400_10LedBlue_Extractor23.eps}
|
---|
542 | \includegraphics[width=0.31\linewidth]{RelArrTime_Pixel400_10LedBlue_Extractor32.eps}
|
---|
543 | \caption{Example of a two distributions of relative arrival times of an outer pixel with respect to
|
---|
544 | the arrival time of the reference pixel Nr. 1. The left plot shows the result using the half-maximum of the spline (extractor \#23), the right plot shows the result obtained with the digital filter
|
---|
545 | (extractor \#32). A
|
---|
546 | medium sized Blue-pulse (10Leds Blue) has been used which saturates the high-gain readout channel.}
|
---|
547 | \label{fig:reltimesouter10ledsblue}
|
---|
548 | \end{figure}
|
---|
549 |
|
---|
550 | %%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
|
---|
551 |
|
---|
552 | \begin{figure}[htp]
|
---|
553 | \centering
|
---|
554 | \includegraphics[width=0.95\linewidth]{UnsuitTimeVsExtractor-5LedsUV-Colour-12.eps}
|
---|
555 | \caption{Reconstructed arrival time resolutions from a typical, not saturating calibration pulse
|
---|
556 | of colour UV, reconstructed with each of the tested arrival time extractors.
|
---|
557 | The first plots shows the time resolutions obtained for the inner pixels, the second one
|
---|
558 | for the outer pixels. Points
|
---|
559 | denote the mean of all not-excluded pixels, the error bars their RMS.}
|
---|
560 | \label{fig:time:5ledsuv}
|
---|
561 | \end{figure}
|
---|
562 |
|
---|
563 | \begin{figure}[htp]
|
---|
564 | \centering
|
---|
565 | \includegraphics[width=0.95\linewidth]{UnsuitTimeVsExtractor-1LedUV-Colour-04.eps}
|
---|
566 | \caption{Reconstructed arrival time resolutions from the lowest intensity calibration pulse
|
---|
567 | of colour UV (carrying a mean number of 4 photo-electrons),
|
---|
568 | reconstructed with each of the tested arrival time extractors.
|
---|
569 | The first plots shows the time resolutions obtained for the inner pixels, the second one
|
---|
570 | for the outer pixels. Points
|
---|
571 | denote the mean of all not-excluded pixels, the error bars their RMS.}
|
---|
572 | \label{fig:time:1leduv}
|
---|
573 | \end{figure}
|
---|
574 |
|
---|
575 | \begin{figure}[htp]
|
---|
576 | \centering
|
---|
577 | \includegraphics[width=0.95\linewidth]{UnsuitTimeVsExtractor-2LedsGreen-Colour-02.eps}
|
---|
578 | \caption{Reconstructed arrival time resolutions from a typical, not saturating calibration pulse
|
---|
579 | of colour Green, reconstructed with each of the tested arrival time extractors.
|
---|
580 | The first plots shows the time resolutions obtained for the inner pixels, the second one
|
---|
581 | for the outer pixels. Points
|
---|
582 | denote the mean of all not-excluded pixels, the error bars their RMS.}
|
---|
583 | \label{fig:time:2ledsgreen}
|
---|
584 | \end{figure}
|
---|
585 |
|
---|
586 | \begin{figure}[htp]
|
---|
587 | \centering
|
---|
588 | \includegraphics[width=0.95\linewidth]{UnsuitTimeVsExtractor-23LedsBlue-Colour-00.eps}
|
---|
589 | \caption{Reconstructed arrival time resolutions from the highest intensity calibration pulse
|
---|
590 | of colour blue, reconstructed with each of the tested arrival time extractors.
|
---|
591 | The first plots shows the time resolutions obtained for the inner pixels, the second one
|
---|
592 | for the outer pixels. Points
|
---|
593 | denote the mean of all not-excluded pixels, the error bars their RMS.}
|
---|
594 | \label{fig:time:23ledsblue}
|
---|
595 | \end{figure}
|
---|
596 |
|
---|
597 | %%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
|
---|
598 |
|
---|
599 | \begin{figure}[htp]
|
---|
600 | \centering
|
---|
601 | \includegraphics[width=0.95\linewidth]{TimeResExtractor-5LedsUV-Colour-12.eps}
|
---|
602 | \caption{Reconstructed arrival time resolutions from a typical, not saturating calibration pulse
|
---|
603 | of colour UV, reconstructed with each of the tested arrival time extractors.
|
---|
604 | The first plots shows the time resolutions obtained for the inner pixels, the second one
|
---|
605 | for the outer pixels. Points
|
---|
606 | denote the mean of all not-excluded pixels, the error bars their RMS.}
|
---|
607 | \label{fig:time:5ledsuv}
|
---|
608 | \end{figure}
|
---|
609 |
|
---|
610 | \begin{figure}[htp]
|
---|
611 | \centering
|
---|
612 | \includegraphics[width=0.95\linewidth]{TimeResExtractor-1LedUV-Colour-04.eps}
|
---|
613 | \caption{Reconstructed arrival time resolutions from the lowest intensity calibration pulse
|
---|
614 | of colour UV (carrying a mean number of 4 photo-electrons),
|
---|
615 | reconstructed with each of the tested arrival time extractors.
|
---|
616 | The first plots shows the time resolutions obtained for the inner pixels, the second one
|
---|
617 | for the outer pixels. Points
|
---|
618 | denote the mean of all not-excluded pixels, the error bars their RMS.}
|
---|
619 | \label{fig:time:1leduv}
|
---|
620 | \end{figure}
|
---|
621 |
|
---|
622 | \begin{figure}[htp]
|
---|
623 | \centering
|
---|
624 | \includegraphics[width=0.95\linewidth]{TimeResExtractor-2LedsGreen-Colour-02.eps}
|
---|
625 | \caption{Reconstructed arrival time resolutions from a typical, not saturating calibration pulse
|
---|
626 | of colour Green, reconstructed with each of the tested arrival time extractors.
|
---|
627 | The first plots shows the time resolutions obtained for the inner pixels, the second one
|
---|
628 | for the outer pixels. Points
|
---|
629 | denote the mean of all not-excluded pixels, the error bars their RMS.}
|
---|
630 | \label{fig:time:2ledsgreen}
|
---|
631 | \end{figure}
|
---|
632 |
|
---|
633 | \begin{figure}[htp]
|
---|
634 | \centering
|
---|
635 | \includegraphics[width=0.95\linewidth]{TimeResExtractor-23LedsBlue-Colour-00.eps}
|
---|
636 | \caption{Reconstructed arrival time resolutions from the highest intensity calibration pulse
|
---|
637 | of colour blue, reconstructed with each of the tested arrival time extractors.
|
---|
638 | The first plots shows the time resolutions obtained for the inner pixels, the second one
|
---|
639 | for the outer pixels. Points
|
---|
640 | denote the mean of all not-excluded pixels, the error bars their RMS.}
|
---|
641 | \label{fig:time:23ledsblue}
|
---|
642 | \end{figure}
|
---|
643 |
|
---|
644 |
|
---|
645 | \begin{figure}[htp]
|
---|
646 | \centering
|
---|
647 | \includegraphics[width=0.95\linewidth]{TimeResVsCharge-Area-21.eps}
|
---|
648 | \caption{Reconstructed mean arrival time resolutions as a function of the extracted mean number of
|
---|
649 | photo-electrons for the weighted sliding window with a window size of 8 FADC slices (extractor \#21).
|
---|
650 | Error bars denote the
|
---|
651 | spread (RMS) of the time resolutions over the investigated channels.
|
---|
652 | The marker colours show the applied
|
---|
653 | pulser colour, except for the last (green) point where all three colours were used.}
|
---|
654 | \label{fig:time:dep20}
|
---|
655 | \end{figure}
|
---|
656 |
|
---|
657 | \begin{figure}[htp]
|
---|
658 | \centering
|
---|
659 | \includegraphics[width=0.95\linewidth]{TimeResVsCharge-Area-24.eps}
|
---|
660 | \caption{Reconstructed mean arrival time resolutions as a function of the extracted mean number of
|
---|
661 | photo-electrons for the half-maximum searching spline (extractor \#23). Error bars denote the
|
---|
662 | spread (RMS) of the time resolutions over the investigated channels.
|
---|
663 | The marker colours show the applied
|
---|
664 | pulser colour, except for the last (green) point where all three colours were used.}
|
---|
665 | \label{fig:time:dep23}
|
---|
666 | \end{figure}
|
---|
667 |
|
---|
668 |
|
---|
669 | \begin{figure}[htp]
|
---|
670 | \centering
|
---|
671 | \includegraphics[width=0.95\linewidth]{TimeResVsCharge-Area-30.eps}
|
---|
672 | \caption{Reconstructed mean arrival time resolutions as a function of the extracted signal
|
---|
673 | for the digital filter with UV weights and 6 slices (extractor \#30). Error bars denote the
|
---|
674 | spread (RMS) of the time resolutions over the investigated channels.
|
---|
675 | The marker colours show the applied
|
---|
676 | pulser colour, except for the last (green) point where all three colours were used.}
|
---|
677 | \label{fig:time:dep30}
|
---|
678 | \end{figure}
|
---|
679 |
|
---|
680 |
|
---|
681 | \begin{figure}[htp]
|
---|
682 | \centering
|
---|
683 | \includegraphics[width=0.95\linewidth]{TimeResVsCharge-Area-31.eps}
|
---|
684 | \caption{Reconstructed mean arrival time resolutions as a function of the extracted signal
|
---|
685 | for the digital filter with UV weights and 4 slices (extractor \#32). Error bars denote the
|
---|
686 | spread (RMS) of the time resolutions over the investigated channels.
|
---|
687 | The marker colours show the applied
|
---|
688 | pulser colour, except for the last (green) point where all three colours were used.}
|
---|
689 | \label{fig:time:dep32}
|
---|
690 | \end{figure}
|
---|
691 |
|
---|
692 |
|
---|
693 |
|
---|
694 |
|
---|
695 |
|
---|
696 | \clearpage
|
---|
697 |
|
---|
698 | \subsection{Pulpo Pulses}
|
---|
699 | \subsection{MC Data}
|
---|
700 | \subsection{Cosmics Data?}
|
---|
701 | The results of this subsection are based on the following runs taken
|
---|
702 | on the 21st of September 2004.
|
---|
703 | \begin{itemize}
|
---|
704 | \item{Run 39000}: OffCrab11 at 19.1 degrees zenith angle and 106.2
|
---|
705 | azimuth.
|
---|
706 | \item{Run 39182}: CrabNebula at 19.0 degrees zenith angle and 106.0 azimuth.
|
---|
707 | \end{itemize}
|
---|
708 |
|
---|
709 | \subsection{Pedestals}
|
---|
710 |
|
---|
711 |
|
---|
712 | %%% Local Variables:
|
---|
713 | %%% mode: latex
|
---|
714 | %%% TeX-master: "MAGIC_signal_reco"
|
---|
715 | %%% TeX-master: "MAGIC_signal_reco."
|
---|
716 | %%% TeX-master: "MAGIC_signal_reco"
|
---|
717 | %%% TeX-master: "MAGIC_signal_reco"
|
---|
718 | %%% TeX-master: "MAGIC_signal_reco"
|
---|
719 | %%% TeX-master: "MAGIC_signal_reco"
|
---|
720 | %%% TeX-master: "MAGIC_signal_reco"
|
---|
721 | %%% TeX-master: "MAGIC_signal_reco"
|
---|
722 | %%% TeX-master: "MAGIC_signal_reco"
|
---|
723 | %%% End:
|
---|