Changes between Version 13 and Version 14 of TestMeasurementsClosedLid


Ignore:
Timestamp:
12/09/14 21:39:04 (11 years ago)
Author:
Daniela Dorner
Comment:

added information for bias on

Legend:

Unmodified
Added
Removed
Modified
  • TestMeasurementsClosedLid

    v13 v14  
    2828make sure that the lid is closed
    2929
     30=== bias
     31
     32you need this only, if you want to take data with closed lid and '''bias on''':
     33
     34in case feedback is not yet calibrated: [[BR]]
     35in feedback: [[BR]]
     36CALIBRATE[[BR]]
     37wait until it is calibrated
     38
     39in feedback[[BR]]
     40START 1.1
     41
     42in biasctrl[[BR]]
     43SET_GLOBAL_DAC 1[[BR]]
     44the feedback system should now ramp the voltage automatically to the needed values
     45
    3046=== configure run
    3147