Version 17 (modified by 10 years ago) ( diff ) | ,
---|
Test Measurements with Closed Lid
as I was asked already several times how to do these measurements, I put here some information on how to do it
goal of such measurements
record events with closed lid to study what type of events occur in that case
bias on or off
the bias might be switched on of off for these measurements
bias on: probably muon events (they look different depending on the zd, also the rate changes)
bias off: only electronics noise triggering
settings you probably want to change in this test
- sky position (different zd)
- bias on/off
- threshold
- number of events
how to do such a measurement
sky position
in case you want to do it not in park position:
in drivectrl:
MOVE_TO <zd> <az>
(you just need to position the telescope, tracking is not needed)
in case you want to go to zenith, I recommend something like 2° as directly zenith might be difficult for a real telescope ;)
lid
make sure that the lid is closed
bias [only needed for measurements with bias voltage]
in case feedback is not yet calibrated:
in feedback:
CALIBRATE
wait until it is calibrated
in feedback
START 1.1
in biasctrl
SET_GLOBAL_DAC 1
the feedback system should now ramp the voltage automatically to the needed values
configure run
in mcp:
START <timemax> <nummax> <runtype>
e.g. MCP/START -1 100 custom (will take a run of 100 events; run type custom is needed here; with runtype data it seemed not to work)
set trigger settings
in dimctrl:
RATE_CONTROL/STOP
in ftmctrl:
STOP_TRIGGER
ENABLE_TRIGGER yes
SET_TIMEMARKER_DELAY 123
SET_THRESHOLD -1 <threshold>
e.g. SET_THRESHOLD -1 400 (setting the threshold to 400 DAC counts
START_TRIGGER
in fadctrl:
wait until run is finished and file is closed
you may look at the events after few minutes (when it's transferred from newdaq to daq) on https://www.fact-project.org/viewer/
stop a measurement
in case you want to abort a measurement, eg when the weather improves:
in ftmctrl:
STOP_TRIGGER
in mcp
STOP
in fadctrl:
CLOSE_OPEN_FILES