source: trunk/MagicSoft/TDAS-Extractor/Calibration.tex@ 6422

Last change on this file since 6422 was 6419, checked in by gaug, 20 years ago
*** empty log message ***
File size: 39.2 KB
Line 
1
2
3In this section, we describe the tests performed using light pulses of different colour,
4pulse shapes and intensities with the MAGIC LED Calibration Pulser Box \cite{hardware-manual}.
5\par
6The LED pulser system is able to provide fast light pulses of 3--4\,ns FWHM
7with intensities ranging from 3--4 to more than 500 photo-electrons in one inner photo-multiplier of the
8camera. These pulses can be produced in three colours {\textit {\bf green, blue}} and
9{\textit{\bf UV}}.
10
11\begin{table}[htp]
12\centering
13\begin{tabular}{|c|c|c|c|c|c|c|}
14\hline
15\hline
16\multicolumn{7}{|c|}{The possible pulsed light colours} \\
17\hline
18\hline
19Colour & Wavelength & Spectral Width & Min. Nr. & Max. Nr. & Secondary & FWHM \\
20 & [nm] & [nm] & Phe's & Phe's & Pulses & Pulse [ns]\\
21\hline
22Green & 520 & 40 & 6 & 120 & yes & 3--4 \\
23\hline
24Blue & 460 & 30 & 6 & 500 & yes & 3--4 \\
25\hline
26UV & 375 & 12 & 3 & 50 & no & 2--3 \\
27\hline
28\hline
29\end{tabular}
30\caption{The pulser colours available from the calibration system}
31\label{tab:pulsercolours}
32\end{table}
33
34Table~\ref{tab:pulsercolours} lists the available colours and intensities and
35figures~\ref{fig:pulseexample1leduv} and~\ref{fig:pulseexample23ledblue} show exemplary pulses
36as registered by the FADCs.
37Whereas the UV-pulse is very stable, the green and blue pulses show sometimes smaller secondary
38pulses after about 10--40\,ns from the main pulse.
39One can see that the very stable UV-pulses are unfortunately only available in such intensities as to
40not saturate the high-gain readout channel. However, the brightest combination of light pulses easily
41saturates all channels in the camera, but does not reach a saturation of the low-gain readout.
42\par
43Our tests can be classified into three subsections:
44
45\begin{enumerate}
46\item Un-calibrated pixels and events: These tests measure the percentage of failures of the extractor
47resulting either in a pixel declared as un-calibrated or in an event which produces a signal ouside
48of the expected Gaussian distribution.
49\item Number of photo-electrons: These tests measure the reconstructed numbers of photo-electrons, their
50spread over the camera and the ratio of the obtained mean values for outer and inner pixels, respectively.
51\item Linearity tests: These tests measure the linearity of the extractor with respect to pulses of
52different intensity and colour.
53\item Time resolution: These tests show the time resolution and stability obtained with different
54intensities and colours.
55\end{enumerate}
56
57\begin{figure}[htp]
58\centering
59\includegraphics[width=0.48\linewidth]{1LedUV_Pulse_Inner.eps}
60\includegraphics[width=0.48\linewidth]{1LedUV_Pulse_Outer.eps}
61\caption{Example of a calibration pulse from the lowest available intensity (1\,Led UV).
62The left plot shows the signal obtained in an inner pixel, the right one the signal in an outer pixel.
63Note that the pulse height fluctuates much more than suggested from these pictures. Especially, a
64zero-pulse is also possible.}
65\label{fig:pulseexample1leduv}
66\end{figure}
67
68\begin{figure}[htp]
69\centering
70\includegraphics[width=0.48\linewidth]{23LedsBlue_Pulse_Inner.eps}
71\includegraphics[width=0.48\linewidth]{23LedsBlue_Pulse_Outer.eps}
72\caption{Example of a calibration pulse from the highest available mono-chromatic intensity (23\,Leds Blue).
73The left plot shows the signal obtained in an inner pixel, the right one the signal in an outer pixel.
74One the left side of both plots, the (saturated) high-gain channel is visible,
75on the right side from FADC slice 18 on,
76the delayed low-gain
77pulse appears. Note that in the left plot, there is a secondary pulses visible in the tail of the
78high-gain pulse. }
79\label{fig:pulseexample23ledblue}
80\end{figure}
81
82We used data taken on the 7$^{th}$ of June, 2004 with different pulser LED combinations, each taken with
8316384 events. 19 different calibration configurations have been tested.
84The corresponding MAGIC data run numbers range from nr. 31741 to 31772. These data was taken
85before the latest camera repair access which resulted in a replacement of about 2\% of the pixels known to be
86mal-functionning at that time.
87There is thus a lower limit to the number of un-calibrated pixels of about 1.5--2\% of known
88mal-functionning photo-multipliers.
89\par
90Although we had looked at and tested all colour and extractor combinations resulting from these data,
91we refrain ourselves to show here only exemplary behaviour and results of extractors.
92All plots, including those which are not displayed in this TDAS, can be retrieved from the following
93locations:
94
95\begin{verbatim}
96http://www.magic.ifae.es/~markus/pheplots/
97http://www.magic.ifae.es/~markus/timeplots/
98\end{verbatim}
99
100%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
101
102\subsection{Un-Calibrated Pixels and Events}
103
104The MAGIC calibration software incorporates a series of checks to sort out mal-functionning pixels.
105Except for the software bug searching criteria, the following exclusion criteria can apply:
106
107\begin{enumerate}
108\item The reconstructed mean signal is less than 2.5 times the extractor resolution $R$ from zero.
109(2.5 Pedestal RMS in the case of the simple fixed window extractors, see section~\ref{sec:pedestals}).
110This criterium essentially cuts out
111dead pixels.
112\item The reconstructed mean signal error is smaller than its value. This criterium cuts out
113signal distributions which fluctuate so much that their RMS is bigger than its mean value. This
114criterium cuts out ``ringing'' pixels or mal-functionning extractors.
115\item The reconstructed mean number of photo-electrons lies 4.5 sigma outside
116the distribution of photo-electrons obtained with the inner or outer pixels in the camera, respectively.
117This criterium cuts out pixels channels with apparently deviating (hardware) behaviour compared to
118the rest of the camera readout\footnote{This criteria is not applied any more in the standard analysis,
119although here, we kept using it}.
120\item All pixels with reconstructed negative mean signal or with a
121mean numbers of photo-electrons smaller than one. Pixels with a negative pedestal RMS subtracted
122sigma occur, especially when stars are focussed onto that pixel during the pedestal taking (resulting
123in a large pedestal RMS), but have moved to another pixel during the calibration run. In this case, the
124number of photo-electrons would result artificially negative. If these
125channels do not show any other deviating behaviour, their number of photo-electrons gets replaced by the
126mean number of photo-electrons in the camera, and the channel is further calibrated as normal.
127\end{enumerate}
128
129Moreover, the number of events are counted which have been reconstructed outside a 5 sigma region
130from the mean signal. These events are called ``outliers''. Figure~\ref{fig:outlier} shows a typical
131outlier obtained with the digital filter applied to a low-gain signal and figure~\ref{fig:unsuited:all}
132shows the average number of all excluded pixels and outliers obtained from all 19 calibration configurations.
133One can already see that the largest window sizes yield a high number of un-calibrated pixels, mostly
134due to the missing ability to recognize the low-intensity pulses (see later). One can also see that
135the amplitude extracting spline yields a higher number of outliers than the rest of the extractors.
136The global champion in lowest number of un-calibrated pixels results to be
137{\textit{\bf MExtractTimeAndChargeDigitalFilter}} with the correct calibration weights over 4 FADC slices
138(extractor \#31). The one with the lowest number of outliers is
139{\textit{\bf MExtractFixedWindowPeakSearch}} with an extraction range of 2 slices (extractor \#11).
140
141\begin{figure}[htp]
142\centering
143\includegraphics[width=0.95\linewidth]{Outlier.eps}
144\caption{Example of an event classified as ``un-calibrated''. The histogram has been obtained
145using the digital filter (extractor \#32) applied to a high-intensity blue pulse (run 31772).
146The event marked as ``outlier'' clearly has been mis-reconstructed. It lies outside the 5 sigma
147region from the fitted mean.}
148\label{fig:outlier}
149\end{figure}
150
151\begin{figure}[htp]
152\centering
153\includegraphics[height=0.75\textheight]{UnsuitVsExtractor-all.eps}
154\caption{Uncalibrated pixels and pixels outside of the Gaussian distribution averaged over all available
155calibration runs.}
156\label{fig:unsuited:all}
157\end{figure}
158
159The following figures~\ref{fig:unsuited:5ledsuv},~\ref{fig:unsuited:1leduv},~\ref{fig:unsuited:2ledsgreen}
160and~\ref{fig:unsuited:23ledsblue} show the resulting numbers of un-calibrated pixels and events for
161different colours and intensities. Because there is a strong anti-correlation between the number of
162excluded channels and the number of outliers per event, we have chosen to show these numbers together.
163
164\par
165
166\begin{figure}[htp]
167\centering
168\includegraphics[height=0.95\textheight]{UnsuitVsExtractor-5LedsUV-Colour-13.eps}
169\caption{Uncalibrated pixels and pixels outside of the Gaussian distribution for a typical calibration
170pulse of UV-light which does not saturate the high-gain readout.}
171\label{fig:unsuited:5ledsuv}
172\end{figure}
173
174\begin{figure}[htp]
175\centering
176\includegraphics[height=0.95\textheight]{UnsuitVsExtractor-1LedUV-Colour-04.eps}
177\caption{Uncalibrated pixels and pixels outside of the Gaussian distribution for a very low
178intensity pulse.}
179\label{fig:unsuited:1leduv}
180\end{figure}
181
182\begin{figure}[htp]
183\centering
184\includegraphics[height=0.95\textheight]{UnsuitVsExtractor-2LedsGreen-Colour-02.eps}
185\caption{Uncalibrated pixels and pixels outside of the Gaussian distribution for a typical green pulse.}
186\label{fig:unsuited:2ledsgreen}
187\end{figure}
188
189\begin{figure}[htp]
190\centering
191\includegraphics[height=0.95\textheight]{UnsuitVsExtractor-23LedsBlue-Colour-00.eps}
192\caption{Uncalibrated pixels and pixels outside of the Gaussian distribution for a high-intensity blue pulse.}
193\label{fig:unsuited:23ledsblue}
194\end{figure}
195
196One can see that in general, big extraction windows raise the
197number of un-calibrated pixels and are thus less stable. Especially for the very low-intensity
198\textit{\bf 1Led\,UV}-pulse, the big extraction windows summing 8 or more slices, cannot calibrate more
199than 50\%
200of the inner pixels (fig.~\ref{fig:unsuited:1leduv}). This is an expected behavior since big windows
201add up more noise which in turn makes the search for the small signal more difficult.
202\par
203In general, one can also find that all ``sliding window''-algorithms (extractors \#17-32) discard
204less pixels than the corresponding ``fixed window''-ones (extractors \#1--16). The digital filter with
205the correct weights (extractors \#30-33) discards the least number of pixels and is also robust against
206slight modifications of its weights (extractors \#28--30). The robustness gets lost when the high-gain and
207low-gain weights are inverted (extractors \#31--39, see fig.~\ref{fig:unsuited:23ledsblue}).
208\par
209Also the ``spline'' algorithms on small
210windows (extractors \#23--25) discard less pixels than the previous extractors.
211\par
212It seems also that the spline algorithm extracting the amplitude of the signal produces an over-proportional
213number of excluded events in the low-gain. The same, however in a less significant manner, holds for
214the digital filter with high-low-gain inverted weights. The limit of stability with respect to
215changes in the pulse form seems to be reached, there.
216\par
217Concerning the numbers of outliers, one can conclude that in general, the numbers are very low never exceeding
2180.1\% except for the ampltiude-extracting spline which seems to mis-reconstruct a certain type of events.
219\par
220In conclusion, already this first test excludes all extractors with too large window sizes because
221they are not able to extract cleanly small signals produced by about 4 photo-electrons. Moreover,
222some extractors do not reproduce the signals as expected in the low-gain.
223
224%The excluded extractors are:
225%\begin{itemize}
226%\item: MExtractFixedWindow Nr. 3--5
227%\item: MExtractFixedWindowSpline Nr. 6--11 (all)
228%\item: MExtractFixedWindowPeakSearch Nr. 14--16
229%\item: MExtractTimeAndChargeSlidingWindow Nr. 21--22
230%\item: MExtractTimeAndChargeSpline Nr. 23 and 27
231%\end{itemize}
232
233\clearpage
234
235%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
236
237\subsection{Number of Photo-Electrons \label{sec:photo-electrons}}
238
239Assuming that the readout chain adds only negligible noise to the one
240introduced by the photo-multiplier itself, one can make the assumption that the variance of the
241true signal $S$ is the amplified Poisson variance of the number of photo-electrons,
242multiplied with the excess noise of the photo-multiplier which itself is
243characterized by the excess-noise factor $F$.
244
245\begin{equation}
246Var(S) = F^2 \cdot Var(N_{phe}) \cdot \frac{<S>^2}{<N_{phe}>^2}
247\label{eq:excessnoise}
248\end{equation}
249
250After introducing the effect of the night-sky background (eq.~\ref{eq:rmssubtraction})
251in formula~\ref{eq:excessnoise} and assuming that the variance of the number of photo-electrons is equal
252to the mean number of photo-electrons (because of the Poisson distribution),
253one obtains an expression to retrieve the mean number of photo-electrons impinging on the pixel from the
254mean extracted signal $<\widehat{S}>$,
255its variance $Var(\widehat{S})$ and the RMS of the extracted signal obtained from
256pure pedestal runs $R$ (see section~\ref{sec:ffactor}):
257
258\begin{equation}
259<N_{phe}> \approx F^2 \cdot \frac{<\widehat{S}>^2}{Var(\widehat{S}) - R^2}
260\label{eq:pheffactor}
261\end{equation}
262
263In theory, eq.~\ref{eq:pheffactor} must not depend on the extractor! Effectively, we will use it to test the
264quality of our extractors by requiring that a valid extractor yields the same number of photo-electrons
265for all pixels of a same type and does not deviate from the number obtained with other extractors.
266As the camera is flat-fielded, but the number of photo-electrons impinging on an inner and an outer pixel is
267different, we also use the ratio of the mean numbers of photo-electrons from the outer pixels to the one
268obtained from the inner pixels as a test variable. In the ideal case, it should always yield its central
269value of about 2.6$\pm$0.1~\cite{michele-diploma}.
270\par
271In our case, there is an additional complication due to the fact that the green and blue coloured light pulses
272show secondary pulses which destroy the Poisson behaviour of the number of photo-electrons. We will
273have to split our sample of extractors into those being affected by the secondary pulses and those
274being immune to this effect.
275\par
276Figures~\ref{fig:phe:5ledsuv},~\ref{fig:phe:1leduv},~\ref{fig:phe:2ledsgreen}~and~\ref{fig:phe:23ledsblue} show
277some of the obtained results. Although one can see a rather good stability for the standard
278{\textit{\bf 5\,Leds\,UV}}\ pulse, except for the extractors {\textit{\bf MExtractFixedWindowPeakSearch}}, initialized
279with an extraction window of 2 slices and {\textit{\bf MExtractTimeAndChargeDigitalFilter}}, initialized with
280an extraction window of 4 slices (extractor \#29).
281\par
282There is a considerable difference for all shown non-standard pulses. Especially the pulses from green
283and blue LEDs
284show a clear dependency of the number of photo-electrons on the extraction window. Only the largest
285extraction windows seem to catch the entire range of (jittering) secondary pulses and get the ratio
286of outer vs. inner pixels right. However, they (obviously) over-estimate the number of photo-electrons
287in the primary pulse.
288\par
289The strongest discrepancy is observed in the low-gain extraction (fig.~\ref{fig:phe:23ledsblue}) where all
290fixed window extractors with too small extraction windows fail to reconstruct the correct numbers.
291This has to do with the fact that
292the fixed window extractors fail to do catch a significant part of the (larger) pulse because of the
2931~FADC slice event-to-event jitter.
294
295
296\begin{figure}[htp]
297\centering
298\includegraphics[height=0.92\textheight]{PheVsExtractor-5LedsUV-Colour-13.eps}
299\caption{Number of photo-electrons from a typical, not saturating calibration pulse of colour UV,
300reconstructed with each of the tested signal extractors.
301The first plots shows the number of photo-electrons obtained for the inner pixels, the second one
302for the outer pixels and the third shows the ratio of the mean number of photo-electrons for the
303outer pixels divided by the mean number of photo-electrons for the inner pixels. Points
304denote the mean of all not-excluded pixels, the error bars their RMS.}
305\label{fig:phe:5ledsuv}
306\end{figure}
307
308\begin{figure}[htp]
309\centering
310\includegraphics[height=0.92\textheight]{PheVsExtractor-1LedUV-Colour-04.eps}
311\caption{Number of photo-electrons from a typical, very low-intensity calibration pulse of colour UV,
312reconstructed with each of the tested signal extractors.
313The first plots shows the number of photo-electrons obtained for the inner pixels, the second one
314for the outer pixels and the third shows the ratio of the mean number of photo-electrons for the
315outer pixels divided by the mean number of photo-electrons for the inner pixels. Points
316denote the mean of all not-excluded pixels, the error bars their RMS.}
317\label{fig:phe:1leduv}
318\end{figure}
319
320\begin{figure}[htp]
321\centering
322\includegraphics[height=0.92\textheight]{PheVsExtractor-2LedsGreen-Colour-02.eps}
323\caption{Number of photo-electrons from a typical, not saturating calibration pulse of colour green,
324reconstructed with each of the tested signal extractors.
325The first plots shows the number of photo-electrons obtained for the inner pixels, the second one
326for the outer pixels and the third shows the ratio of the mean number of photo-electrons for the
327outer pixels divided by the mean number of photo-electrons for the inner pixels. Points
328denote the mean of all not-excluded pixels, the error bars their RMS.}
329\label{fig:phe:2ledsgreen}
330\end{figure}
331
332
333\begin{figure}[htp]
334\centering
335\includegraphics[height=0.92\textheight]{PheVsExtractor-23LedsBlue-Colour-00.eps}
336\caption{Number of photo-electrons from a typical, high-gain saturating calibration pulse of colour blue,
337reconstructed with each of the tested signal extractors.
338The first plots shows the number of photo-electrons obtained for the inner pixels, the second one
339for the outer pixels and the third shows the ratio of the mean number of photo-electrons for the
340outer pixels divided by the mean number of photo-electrons for the inner pixels. Points
341denote the mean of all not-excluded pixels, the error bars their RMS.}
342\label{fig:phe:23ledsblue}
343\end{figure}
344
345One can see that all extractors using a large window belong to the class of extractors being affected
346by the secondary pulses, except for the digital filter. The only exception to this rule is the digital filter
347which - despite of its 6 slices extraction window - seems to filter out all the secondary pulses.
348\par
349The extractor {\textit{\bf MExtractFixedWindowPeakSearch}} at low extraction windows apparently yields chronically low
350numbers of photo-electrons. This is due to the fact that the decision to fix the extraction window is
351made sometimes by an inner pixel and sometimes by an outer one since the camera is flat-fielded and the
352pixel carrying the largest non-saturated peak-search window is more or less found by a random signal
353fluctuation. However, inner and outer pixels have a systematic offset of about 0.5 to 1 FADC slices.
354Thus, the extraction fluctuates artificially for one given channel which results in a systematically
355large variance and thus in a systematically low reconstructed number of photo-electrons. This test thus
356excludes the extractors \#11--13.
357\par
358Moreover, one can see that the extractors applying a small fixed window do not get the ratio of
359photo-electrons correctly between outer to inner pixels for the green and blue pulses.
360\par
361The extractor {\textit{\bf MExtractTimeAndChargeDigitalFilter}} seems to be stable against modifications in the
362exact form of the weights in the high-gain readout channel since all applied weights yield about
363the same number of photo-electrons and the same ratio of outer vs. inner pixels. This statement does not
364hold any more for the low-gain, as can be seen in figure~\ref{fig:phe:23ledsblue}. There, the application
365of high-gain weights to the low-gain signal (extractors \#34--39) produces a too low number of photo-electrons
366and also a too low ratio of outer vs. inner pixels.
367\par
368All sliding window and spline algorithms yield a stable ratio of outer vs. inner pixels in the low-gain,
369however the effect of raising the number of photo-electrons with the extraction window is very pronounced.
370Note that in figure~\ref{fig:phe:23ledsblue}, the number of photo-electrons rises by about a factor 1.4,
371which is slightly higher than in the case of the high-gain channel (figure~\ref{fig:phe:2ledsgreen}).
372\par
373Concluding, there is no fixed window extractor yielding the correct number of photo-electrons
374for the low-gain, except for the largest extraction window of 8 and 10 low-gain slices.
375Either the number of photo-electrons itself is wrong or the ratio of outer vs. inner pixels is
376not correct. All sliding window algorithms seem to reproduce the correct numbers if one takes into
377account the after-pulse behaviour of the light pulser itself. The digital filter seems to be
378unstable against exchanging the pulse form to match the slimmer high-gain pulses, though.
379
380\par
381\ldots {\textit{\bf EXCLUDED : CW4, UV4 No stability High-gain vs. LoGain}}
382\par
383
384%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
385
386\subsection{Linearity \label{sec:calibration:linearity}}
387
388\begin{figure}[htp]
389\centering
390\includegraphics[width=0.75\linewidth]{PheVsCharge-3.eps}
391\caption{Conversion factor $c_{phe}$ for two exemplary inner pixels (upper plots)
392and two exemplary outer ones (lower plots) obtained with the extractor
393{\textit{MExtractFixedWindow}} on a window size of 6 high-gain and 6 low-gain slices
394(extractor \#3). }
395\label{fig:linear:phevscharge3}
396\end{figure}
397
398In this section, we test the lineary of the conversion factors FADC counts to photo-electrons:
399
400\begin{equation}
401c_{phe} =\ <Phe> / <\widehat{S}>
402\end{equation}
403
404As the photo-multiplier and the subsequent
405optical transmission devices~\cite{david} is a linear device over a
406wide dynamic range, the number of photo-electrons per charge has to remain constant over the tested
407linearity region. We will show here only examples of extractors which were not already excluded in the
408previous section.
409\par
410A first test concerns the stability of the conversion factor: mean number of averaged photo-electrons
411per FADC counts over the
412tested intensity region. A much more detailed investigation on the linearity will be shwon in a
413separate TDAS~\cite{tdas-calibration}.
414
415\par
416Figure~\ref{fig:linear:phevscharge3} shows the conversion factor $c_{phe}$
417obtained for different light intensities
418and colours for two exemplary inner and two exemplary outer pixels using a fixed window on
4196 FADC slices. One can clearly see the difference
420between the high-gain ($<$100\ phes) and the low-gain ($>$100\ phes) region and
421a rather good stability of $c_{phe}$ for each region separately, except for the highest intensities
422($>$400\ phes). We conclude
423that the fixed window extractor \#3 is a linear extractor for both high-gain and low-gain regions,
424separately below a signal of about 300 photo-elecrons.
425\par
426
427\begin{figure}[htp]
428\centering
429\includegraphics[width=0.75\linewidth]{PheVsCharge-8.eps}
430\caption{Conversion factor $c_{phe}$ for two exemplary inner pixels (upper plots)
431and two exemplary outer ones (lower plots) obtained with the extractor
432{\textit{MExtractFixedWindowSpline}}
433on a window size of 6 high-gain and 6 low-gain slices (extractor \#8). }
434\label{fig:linear:phevscharge8}
435\end{figure}
436
437Figure~\ref{fig:linear:phevscharge8} shows the conversion factors using an integrated spline over
438a fixed window of 7 FADC slices. There is a rather stability in
439the high-gain region ($<$100\ phes), but the low-gain region fluctuates a lot, especially for the two
440 outer pixels. We conclude that the fixed window spline extractor has a bad linearity
441and is not robust in the low-gain extraction.
442\par
443
444\begin{figure}[htp]
445\centering
446\includegraphics[width=0.75\linewidth]{PheVsCharge-14.eps}
447\caption{Conversion factor $c_{phe}$ for two exemplary inner pixels (upper plots)
448and two exemplary outer ones (lower plots) obtained with the extractor
449{\textit{MExtractFixedWindowPeakSearch}} on a window size of 6 high-gain and 6 low-gain slices
450(extractor \#14). }
451\label{fig:linear:phevscharge14}
452\end{figure}
453
454Figure~\ref{fig:linear:phevscharge14} shows the conversion factors using a fixed window obtained with
455a global peak search over the camera. A similiar result the fixed window is obtained where there is
456stability up to about 300 photo-electrons. We conclude
457that the fixed window peak search extractor \#14 is linear for both high-gain and low-gain regions,
458separately, below a signal of about 300 photo-elecrons.
459\par
460
461
462\begin{figure}[htp]
463\centering
464\includegraphics[width=0.75\linewidth]{PheVsCharge-20.eps}
465\caption{Example of a the development of the conversion factor FADC counts to photo-electrons for two
466exemplary inner pixels (upper plots) and two exemplary outer ones (lower plots) obtained with the extractor
467{\textit{MExtractTimeAndChargeSlidingWindow}}
468on a window size of 6 high-gain and 6 low-gain slices (extractor \#20). }
469\label{fig:linear:phevscharge20}
470\end{figure}
471
472Figure~\ref{fig:linear:phevscharge20} shows the conversion factors using a sliding fixed window.
473A much higher dynamic range is obtained mainting stability up to further than 500 photo-electrons.
474\par
475
476
477
478\begin{figure}[htp]
479\centering
480\includegraphics[width=0.75\linewidth]{PheVsCharge-25.eps}
481\caption{Conversion factor $c_{phe}$ for two exemplary inner pixels (upper plots)
482and two exemplary outer ones (lower plots) obtained with the extractor
483{\textit{MExtractTimeAndChargeSpline}} with window size of 2 high-gain and 3 low-gain slices
484(extractor \#25). }
485\label{fig:linear:phevscharge25}
486\end{figure}
487
488Figure~\ref{fig:linear:phevscharge25} shows the conversion factors using a sliding spline
489extractor with an integration window of 2 FADC slices in the high-gain and 3 FADC slices in the
490low-gain. The increase of integration window in the low-gain seems to lead to an systematic
491increase in the conversion factor above 200 photo-electrons. If one uses this extractor, probably this
492effect will have to be corrected for.
493
494\par
495
496\begin{figure}[htp]
497\centering
498\includegraphics[width=0.75\linewidth]{PheVsCharge-30.eps}
499\caption{Conversion factor $c_{phe}$ for two exemplary inner pixels (upper plots)
500and two exemplary outer ones (lower plots) obtained with the extractor
501{\textit{MExtractTimeAndChargeDigitalFilter}}
502using a window size of 6 high-gain and 6 low-gain slices with UV-weights (extractor \#30). }
503\label{fig:linear:phevscharge30}
504\end{figure}
505
506Figure~\ref{fig:linear:phevscharge30} shows the conversion factors using a digital filter applied on 6 FADC slices with weights calculated from
507the UV-calibration pulse.
508One can see that all calibration blue and green calibration pulses at low and intermediate intensity fall
509 out of the linear region, moreover there seems to be
510a systematic offset between high-gain and low-gain. These offsets have to corrected for in any way, however the loss of stability against the
511exact pulse form in the high-gain is more problematic.
512
513\par
514
515\begin{figure}[htp]
516\centering
517\includegraphics[width=0.75\linewidth]{PheVsCharge-31.eps}
518\caption{Conversion factor $c_{phe}$ for two exemplary inner pixels (upper plots)
519and two exemplary outer ones (lower plots) obtained with the extractor
520{\textit{MExtractTimeAndChargeDigitalFilter}} using a window size of
5214 high-gain and 4 low-gain slices (extractor \#31). }
522\label{fig:linear:phevscharge31}
523\end{figure}
524
525\clearpage
526
527\subsection{Time Resolution}
528
529The extractors \#17--32 are able to extract also the arrival time of each pulse. The calibration
530delivers a fast-rising pulse, uniform over the camera in signal size and time.
531We estimate the time-uniformity to better
532than 300\,ps, a limit due to the different travel times of the light between inner and outer parts of the
533camera. Since the calibraion does not permit a precise measurement of the absolute arrival time, we measure
534the relative arrival time for every channel with respect to a reference channel (usually pixel Nr.\,1):
535
536\begin{equation}
537\delta t_i = t_i - t_1
538\end{equation}
539
540where $t_i$ denotes the reconstructed arrival time of pixel number $i$ and $t_1$ the reconstructed
541arrival time of the reference pixel nr. 1 (software numbering). For one calibration run, one can then fill
542histograms of $\delta t_i$ for each pixel and fit them to the expected Gaussian distribution. The fits
543yield a mean $\mu(\delta t_i)$, comparable to
544systematic offsets in the signal delay, and a sigma $\sigma(\delta t_i)$, a measure of the
545combined time resolutions of pixel $i$ and pixel 1. Assuming that the PMTs and readout channels are
546of a same kind, we obtain an approximate absolute time resolution of pixel $i$ by:
547
548\begin{equation}
549t^{res}_i \approx \sigma(\delta t_i)/sqrt(2)
550\end{equation}
551
552Figures~\ref{fig:reltimesinner10leduv} and~\ref{fig:reltimesouter10leduv} show distributions of $\delta t_i$
553for
554one typical inner pixel and one typical outer pixel and a non-saturating calibration pulse of UV-light,
555obtained with three different extractors. One can see that the first two yield a Gaussian distribution
556to a good approximation, whereas the third extractor shows a three-peak structure and cannot be fitted.
557We discarded that particular extractor for this reason.
558
559\begin{figure}[htp]
560\centering
561\includegraphics[width=0.3\linewidth]{RelArrTime_Pixel97_10LedUV_Extractor32.eps}
562\includegraphics[width=0.32\linewidth]{RelArrTime_Pixel97_10LedUV_Extractor23.eps}
563\includegraphics[width=0.32\linewidth]{RelArrTime_Pixel97_10LedUV_Extractor17.eps}
564\caption{Example of a two distributions of relative arrival times of an inner pixel with respect to
565the arrival time of the reference pixel Nr. 1. The left plot shows the result using the digital filter
566 (extractor \#32), the central plot shows the result obtained with the half-maximum of the spline and the
567right plot the result of the sliding window with a window size of 2 FADC slices (extractor \#17). A
568medium sized UV-pulse (10Leds UV) has been used which does not saturate the high-gain readout channel.}
569\label{fig:reltimesinner10leduv}
570\end{figure}
571
572\begin{figure}[htp]
573\centering
574\includegraphics[width=0.31\linewidth]{RelArrTime_Pixel400_10LedUV_Extractor32.eps}
575\includegraphics[width=0.31\linewidth]{RelArrTime_Pixel400_10LedUV_Extractor23.eps}
576\includegraphics[width=0.31\linewidth]{RelArrTime_Pixel400_10LedUV_Extractor17.eps}
577\caption{Example of a two distributions of relative arrival times of an outer pixel with respect to
578the arrival time of the reference pixel Nr. 1. The left plot shows the result using the digital filter
579 (extractor \#32), the central plot shows the result obtained with the half-maximum of the spline and the
580right plot the result of the sliding window with a window size of 2 FADC slices (extractor \#17). A
581medium sized UV-pulse (10Leds UV) has been used which does not saturate the high-gain readout channel.}
582\label{fig:reltimesouter10leduv}
583\end{figure}
584
585Figures~\ref{fig:reltimesinner10ledsblue} and~\ref{fig:reltimesouter10ledsblue} show distributions of
586$<\delta t_i>$ for
587one typical inner and one typical outer pixel and a high-gain-saturating calibration pulse of blue-light,
588obtained with two different extractors. One can see that the first (extractor \#23) yields a Gaussian
589distribution to a good approximation.
590
591\begin{figure}[htp]
592\centering
593\includegraphics[width=0.31\linewidth]{RelArrTime_Pixel97_10LedBlue_Extractor23.eps}
594\includegraphics[width=0.31\linewidth]{RelArrTime_Pixel97_10LedBlue_Extractor32.eps}
595\caption{Example of a two distributions of relative arrival times of an inner pixel with respect to
596the arrival time of the reference pixel Nr. 1. The left plot shows the result using the half-maximum of the spline (extractor \#23), the right plot shows the result obtained with the digital filter
597(extractor \#32). A
598medium sized Blue-pulse (10Leds Blue) has been used which saturates the high-gain readout channel.}
599\label{fig:reltimesinner10ledsblue}
600\end{figure}
601
602
603
604\begin{figure}[htp]
605\centering
606\includegraphics[width=0.31\linewidth]{RelArrTime_Pixel400_10LedBlue_Extractor23.eps}
607\includegraphics[width=0.31\linewidth]{RelArrTime_Pixel400_10LedBlue_Extractor32.eps}
608\caption{Example of a two distributions of relative arrival times of an outer pixel with respect to
609the arrival time of the reference pixel Nr. 1. The left plot shows the result using the half-maximum of the spline (extractor \#23), the right plot shows the result obtained with the digital filter
610(extractor \#32). A
611medium sized Blue-pulse (10Leds Blue) has been used which saturates the high-gain readout channel.}
612\label{fig:reltimesouter10ledsblue}
613\end{figure}
614
615%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
616
617\begin{figure}[htp]
618\centering
619\includegraphics[width=0.95\linewidth]{UnsuitTimeVsExtractor-5LedsUV-Colour-12.eps}
620\caption{Reconstructed arrival time resolutions from a typical, not saturating calibration pulse
621of colour UV, reconstructed with each of the tested arrival time extractors.
622The first plots shows the time resolutions obtained for the inner pixels, the second one
623for the outer pixels. Points
624denote the mean of all not-excluded pixels, the error bars their RMS.}
625\label{fig:time:5ledsuv}
626\end{figure}
627
628\begin{figure}[htp]
629\centering
630\includegraphics[width=0.95\linewidth]{UnsuitTimeVsExtractor-1LedUV-Colour-04.eps}
631\caption{Reconstructed arrival time resolutions from the lowest intensity calibration pulse
632of colour UV (carrying a mean number of 4 photo-electrons),
633reconstructed with each of the tested arrival time extractors.
634The first plots shows the time resolutions obtained for the inner pixels, the second one
635for the outer pixels. Points
636denote the mean of all not-excluded pixels, the error bars their RMS.}
637\label{fig:time:1leduv}
638\end{figure}
639
640\begin{figure}[htp]
641\centering
642\includegraphics[width=0.95\linewidth]{UnsuitTimeVsExtractor-2LedsGreen-Colour-02.eps}
643\caption{Reconstructed arrival time resolutions from a typical, not saturating calibration pulse
644of colour Green, reconstructed with each of the tested arrival time extractors.
645The first plots shows the time resolutions obtained for the inner pixels, the second one
646for the outer pixels. Points
647denote the mean of all not-excluded pixels, the error bars their RMS.}
648\label{fig:time:2ledsgreen}
649\end{figure}
650
651\begin{figure}[htp]
652\centering
653\includegraphics[width=0.95\linewidth]{UnsuitTimeVsExtractor-23LedsBlue-Colour-00.eps}
654\caption{Reconstructed arrival time resolutions from the highest intensity calibration pulse
655of colour blue, reconstructed with each of the tested arrival time extractors.
656The first plots shows the time resolutions obtained for the inner pixels, the second one
657for the outer pixels. Points
658denote the mean of all not-excluded pixels, the error bars their RMS.}
659\label{fig:time:23ledsblue}
660\end{figure}
661
662%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
663
664\begin{figure}[htp]
665\centering
666\includegraphics[width=0.95\linewidth]{TimeResExtractor-5LedsUV-Colour-12.eps}
667\caption{Reconstructed arrival time resolutions from a typical, not saturating calibration pulse
668of colour UV, reconstructed with each of the tested arrival time extractors.
669The first plots shows the time resolutions obtained for the inner pixels, the second one
670for the outer pixels. Points
671denote the mean of all not-excluded pixels, the error bars their RMS.}
672\label{fig:time:5ledsuv}
673\end{figure}
674
675\begin{figure}[htp]
676\centering
677\includegraphics[width=0.95\linewidth]{TimeResExtractor-1LedUV-Colour-04.eps}
678\caption{Reconstructed arrival time resolutions from the lowest intensity calibration pulse
679of colour UV (carrying a mean number of 4 photo-electrons),
680reconstructed with each of the tested arrival time extractors.
681The first plots shows the time resolutions obtained for the inner pixels, the second one
682for the outer pixels. Points
683denote the mean of all not-excluded pixels, the error bars their RMS.}
684\label{fig:time:1leduv}
685\end{figure}
686
687\begin{figure}[htp]
688\centering
689\includegraphics[width=0.95\linewidth]{TimeResExtractor-2LedsGreen-Colour-02.eps}
690\caption{Reconstructed arrival time resolutions from a typical, not saturating calibration pulse
691of colour Green, reconstructed with each of the tested arrival time extractors.
692The first plots shows the time resolutions obtained for the inner pixels, the second one
693for the outer pixels. Points
694denote the mean of all not-excluded pixels, the error bars their RMS.}
695\label{fig:time:2ledsgreen}
696\end{figure}
697
698\begin{figure}[htp]
699\centering
700\includegraphics[width=0.95\linewidth]{TimeResExtractor-23LedsBlue-Colour-00.eps}
701\caption{Reconstructed arrival time resolutions from the highest intensity calibration pulse
702of colour blue, reconstructed with each of the tested arrival time extractors.
703The first plots shows the time resolutions obtained for the inner pixels, the second one
704for the outer pixels. Points
705denote the mean of all not-excluded pixels, the error bars their RMS.}
706\label{fig:time:23ledsblue}
707\end{figure}
708
709
710\begin{figure}[htp]
711\centering
712\includegraphics[width=0.95\linewidth]{TimeResVsCharge-Area-21.eps}
713\caption{Reconstructed mean arrival time resolutions as a function of the extracted mean number of
714photo-electrons for the weighted sliding window with a window size of 8 FADC slices (extractor \#21).
715Error bars denote the
716spread (RMS) of the time resolutions over the investigated channels.
717The marker colours show the applied
718pulser colour, except for the last (green) point where all three colours were used.}
719\label{fig:time:dep20}
720\end{figure}
721
722\begin{figure}[htp]
723\centering
724\includegraphics[width=0.95\linewidth]{TimeResVsCharge-Area-24.eps}
725\caption{Reconstructed mean arrival time resolutions as a function of the extracted mean number of
726photo-electrons for the half-maximum searching spline (extractor \#23). Error bars denote the
727spread (RMS) of the time resolutions over the investigated channels.
728The marker colours show the applied
729pulser colour, except for the last (green) point where all three colours were used.}
730\label{fig:time:dep23}
731\end{figure}
732
733
734\begin{figure}[htp]
735\centering
736\includegraphics[width=0.95\linewidth]{TimeResVsCharge-Area-30.eps}
737\caption{Reconstructed mean arrival time resolutions as a function of the extracted signal
738for the digital filter with UV weights and 6 slices (extractor \#30). Error bars denote the
739spread (RMS) of the time resolutions over the investigated channels.
740The marker colours show the applied
741pulser colour, except for the last (green) point where all three colours were used.}
742\label{fig:time:dep30}
743\end{figure}
744
745
746\begin{figure}[htp]
747\centering
748\includegraphics[width=0.95\linewidth]{TimeResVsCharge-Area-31.eps}
749\caption{Reconstructed mean arrival time resolutions as a function of the extracted signal
750for the digital filter with UV weights and 4 slices (extractor \#32). Error bars denote the
751spread (RMS) of the time resolutions over the investigated channels.
752The marker colours show the applied
753pulser colour, except for the last (green) point where all three colours were used.}
754\label{fig:time:dep32}
755\end{figure}
756
757%%% Local Variables:
758%%% mode: latex
759%%% TeX-master: "MAGIC_signal_reco"
760%%% End:
Note: See TracBrowser for help on using the repository browser.