Datataking Procedures
First Drs Calibration
Notice: dimscript for this procedure: ScriptsForDimCtrl/FirstDrsCalibration.dim
Remote-datataking: In case of remote data-taking this section can be skipped. In this case you can directly open the lid, and start with the data-taking including the DRS calibration.
- Do _Start_ (DRS Calibration) in the fad-tab
- Take *DRS: Pedestal* (1000 events)
- Take *DRS: Gain* (1000 events)
- Take *Pedestal* (1000 events)
- Switch to FITS file format in the fad-tab
- Take *Pedestal* (1000 events)
- Switch on the temperature control (Feedback-tab: *Start*, *Enable output*)
- Switch on the BIAS voltage (Set the global offset to one DAC count in the Bias tab and wait until the temperature feedback has ramped the voltages properly)
- Take *Pedestal* (1000 events)
- Switch off the bias voltage (*Set all to 0*)
- Open the lid
Normal Datataking
- Move the telescope to wobble position 1
- Make sure Bias voltage is off
- Do _START_ (DRS Calibration) in the fad-tab
- Take *DRS: Pedestal* (1000 events, MCP *Start Run* on GUIs left side)
- Take *DRS: Gain* (1000 events)
- Take *DRS: Pedestal* (1000 events)
- Switch to FITS file format in the fad-tab
- Take *DRS: Pedestal* (1000 events)
- Take *DRS: Time* (1000 events)
- Do _RESET_ (DRS Calibration) to reset the secondary baseline in the fad-tab
- Take *Pedestal* (1000 events)
- Switch to FITS file format in the fad-tab
- Send somebody out at the backdoor to measure the the sky-brightness with the sky-meter at the rough direction the telescope is pointing and write it down in the logbook
- Take *Pedestal* (1000 events)
- Switch Bias on (set to one DAC count and wait for the feedback)
- Take *Pedestal* (1000 events)
- Take *External light pulser* (1000 events)
- Take data with full trigger area (four times 5 minutes)
- Move the telescope to wobble position 2
- Send somebody out at the backdoor to measure the the sky-brightness with the sky-meter at the rough direction the telescope is pointing and write it down in the logbook
- Take *Pedestal* (1000 events)
- Take *External light pulser* (1000 events)
- Take data with full trigger area (four times 5 minutes)
- Go on with the beginning of the data taking procedure
ratescan
- Track a dark region
- low zenith (<30°)
- no bright star
- no direct moonlight in the camera
- examples
- around 0:00 UTC: ra = 13.46667, dec = 27.12833
- around 4:00 UTC: ra = 21.45, dec = 14.9
- Take a run with type ratescan with unlimited event and unlimited duration (for configuring all programms)
- Ratescan/START_THRESHOLD_SCAN x y c
variables:
- x: start threshold
- y: end threshold
- c: stepsize (if negativ ratescan starts with |c| and handle it dynamically, if the ratescan needs too much time)
Configuring manually:
- Stop ratecontrol (I'm not sure if this is necessary)
- ftmctrl/SET_DEAD_TIME 0 (in fact thats the minimum equal to 8ns (you can also set this value in the GUI)
- GUI: no interleaved pedestal, no interleaved extLP, only physic triggers
Ampl. vs Bias measurement
- Track a dark region (same requirements as for the ratescan, see above)
- Do a DRS calibration
- Feedback/Start_Current_Control -1.0 (maybe Enable_Output yes)
- wait for 2 - 3 feedback voltage corrections
- switch to fits format
- Now take an external lp run with 1000 events
- be aware that for x = -1.0, -0.9, its possible to trigger 0 events. Than you can go on with the next overvoltage
- Increase voltage offset by 0.1 V go on with the next light-pulser run
- do this until +0.5 V offset
Write down following in the logbook:
- Offset
- median Voltage
- median Current
- runnumber
Dim Scripts
In FACT++/ScriptsForDimCtrl you will find the following scripts that could do parts of aboves procedures automatically.
script_name | function |
---|---|
FirstDrsCalib.dim | taking a first DRS calibration as described above |
DataTaking1.dim | procedure from step 2 to step 18 |
DataTaking2.dim | procedure from step 21 to step 23 (which btw. is the same as step 16 to step 18) |
ResumeDataTaking.dim | can be used if Datataking crashes during the first of a 4x5min run set to take 4 runs of 5 mins in a row |
Take<source_name>.dim | procedure from step 1 to 23 with for Source <source_name> (2 Wobble positions) |
AmplVsOv.dim | do steps 2 to 5 of the Amplitude vs Bias measurement with an overvoltage of x |
Scripts can be started from a bash:
- dimctrl --batch ScriptsForDimCtrl/<scriptname>.dim
Scripts can be stopped via dimctrl --stop in a bash
if you want to only take the second wobble position the script can be started by: .x ScriptsForDimCtrl/Take<source_name>:2
scripts available for the following sources:
source | <script_name> |
---|---|
Crab | TakeCrab.dim |
Mrk421 | TakeMrk421.dim |
Mrk501 | TakeMrk501.dim |
1ES1218+304 | Take1218.dim |