| 1 | = Trouble Shooting - Hardware |
| 2 | |
| 3 | == Power Cycling the camera |
| 4 | |
| 5 | __Symptom__ |
| 6 | |
| 7 | - A Crate Reset didn't solved a FTU or FAD problem |
| 8 | |
| 9 | __Solution__ |
| 10 | |
| 11 | - stop all scripts: dimctrl --stop from a bash |
| 12 | - set bias to 0: - BIAS_CONTROL/SET_ZERO_VOLTAGE |
| 13 | - Stop Trigger: |
| 14 | - FTM_CONTROL/STOP_TRIGGER |
| 15 | - or Stop Trigger in the left part of the GUI |
| 16 | - FTM_CONTROL/DISCONNECT |
| 17 | - FEEDBACK/STOP |
| 18 | - if fadctrl is in state WritingData |
| 19 | - FAD_CONTROL/CLOSE_OPEN_FILES |
| 20 | - fadctrl now should be in state connected |
| 21 | - FAD_CONTROL/STOP |
| 22 | |
| 23 | Now __power cycle__ the camera |
| 24 | |
| 25 | 0. Switch off the agilent camera in the FACT Container |
| 26 | - open the page 10.0.100.100 from a computer in the internal network of La Palma (can be your computer if you use vpn) |
| 27 | - push 'Camera Off' to stop the camera (now only 'Bias Power' and '24VDC Interlock Power' should be 'yes') |
| 28 | 0. wait for about 15 min |
| 29 | 0. switch the agilent on |
| 30 | - open the page 10.0.100.100 from a computer in the internal network of La Palma |
| 31 | - push 'Camera On' to start the camera (now all 5 points should be 'yes') |
| 32 | 0. make sure that clock-conditioner is locked |
| 33 | 0. follow the [[preparation#startup Start-Up procedure]] to get the system running again |
| 34 | |
| 35 | To check if the __clock-conditioner__ is locked, you may check in the gui in the tab |
| 36 | |
| 37 | - 'Trigger' the led next to the pulldown 'DRS sampling frequency' |
| 38 | - 'FAD' with mouse over on the LED next to 'Reference Clock' whether all 40 values are roughly 1/1024 |
| 39 | (some more information in a post by Patrick: https://www.fact-project.org/logbook/showthread.php?tid=1102&pid=6070#pid6070 ) |
| 40 | |
| 41 | == Switching off the bias crate |
| 42 | In case you need to switch off the bias crate, please do first BIAS_CONTROL/DISCONNECT |